Acta Optica Sinica, Volume. 41, Issue 24, 2411001(2021)

Double-Calibration Method of Imaging Polarimeters

Zijian Han1,2,3、* and Deqing Ren4
Author Affiliations
  • 1Nanjing Institute of Astronomical Optics & Technology, National Astronomical Observatories, Chinese Academy of Sciences, Nanjing, Jiangsu 210042, China;
  • 2CAS Key Laboratory of Astronomical Optics & Technology, Nanjing Institute of Astronomical Optics & Technology, Nanjing, Jiangsu 210042, China;
  • 3University of Chinese Academy of Sciences, Beijing 100049, China
  • 4Physics & Astronomy Department, California State University Northridge, Northridge, Los Angeles 91330-8268, USA;
  • show less
    Figures & Tables(8)
    Schematic diagram of the imaging polarimeter
    Flow chart of the double-calibration process of the imaging polarimeter
    Experimental setup of the imaging polarimeter
    Change curve of the evaluation function value during the optimization process
    Measurement results of the normalized Stokes components. (a) Q/I component; (b) U/I component; (c) V/I component
    • Table 1. Type and key parameters of the components in the imaging polarimeter

      View table

      Table 1. Type and key parameters of the components in the imaging polarimeter

      No.ComponentManufacturer /TypeKey parameter
      1LASThorlabs HNL050R5 mW power, random polarization
      2PThorlabs LPVIS100extinction ratio > 100000∶1 at 633 nm
      3QWPThorlabs WPQ10ME-633retardance accuracy <λ/100 at 633 nm
      4LCVRThorlabs LCC1223-Anon-compensated, retardance accuracy <λ/50
      5WPThorlabs WPQ10separation angle 1°at 633 nmextinction ratio>10000∶1 at 633 nm
      6step motor rotator(mounting with P/QWP)Thorlabs K10CR1absolute accuracy ±0.014°
    • Table 2. Modulation strategy of the imaging polarimeterunit:(°)

      View table

      Table 2. Modulation strategy of the imaging polarimeterunit:(°)

      ElementFast axisazimuthLCVR retardance
      Step 1Step 2Step 3Step 4
      LCVR122.5356.41298.90103.00107.00
      LCVR245.0343.70134.0049.30156.00
    • Table 3. Measurement precisions of the normalized Stokes components before and after the double-calibration

      View table

      Table 3. Measurement precisions of the normalized Stokes components before and after the double-calibration

      NormalizedStokescomponentRMSE
      Before double-calibrationAfter double-calibration
      Q/I2.6970×10-32.1814×10-3
      U/I7.5149×10-33.8351×10-3
      V/I6.5679×10-36.2961×10-3
    Tools

    Get Citation

    Copy Citation Text

    Zijian Han, Deqing Ren. Double-Calibration Method of Imaging Polarimeters[J]. Acta Optica Sinica, 2021, 41(24): 2411001