Acta Photonica Sinica, Volume. 54, Issue 5, 0512002(2025)

Calibration Method of Instrument Matrix for Polarization Metagrating

Zeyu ZHENG, Chao YE, Yucong ZHOU, Chunlian ZHAN, and Han GAO*
Author Affiliations
  • College of Optical and Electronic Technology,China Jiliang University,Hangzhou 310018,China
  • show less
    Figures & Tables(8)
    Schematic diagram of the working principle of the two-dimensional polarization metagrating
    Polarization generation and detection optical path
    1 550 nm linear polarization light detection results and errors
    1 550 nm elliptically polarized light detection results and errors
    1 550 nm partial polarization light detection results and errors
    500 nm fully polarized light detection results
    • Table 1. Root mean square error of 1 550 nm fully polarized light and partially polarized light detection results

      View table
      View in Article

      Table 1. Root mean square error of 1 550 nm fully polarized light and partially polarized light detection results

      Calibration

      method and measurement

      parameters

      Completely polarized lightPartially polarized light
      Linearly polarized lightElliptically polarized light

      AOP

      RMSE/(°)

      DOP

      RMSE

      AOP

      RMSE/(°)

      EOP

      RMSE/(°)

      DOP

      RMSE

      DOP

      RMSE

      Two-step method0.5230.0173.0083.3750.0300.018
      One-step method1.3240.0730.3550.9880.0140.013
      RMSE difference0.8010.0562.6532.3870.0160.005
    • Table 2. Root mean square error of 500 nm fully polarized light and partially polarized light detection results

      View table
      View in Article

      Table 2. Root mean square error of 500 nm fully polarized light and partially polarized light detection results

      Calibration

      method and measurement

      parameters

      Completely polarized lightPartially polarized light
      Linearly polarized lightElliptically polarized light

      AOP

      RMSE/(°)

      DOP

      RMSE

      AOP

      RMSE/(°)

      EOP

      RMSE/(°)

      DOP

      RMSE

      DOP

      RMSE

      Two-step method0.7980.0172.9672.8430.0390.032
      One-step method1.0310.0450.6500.9600.0190.030
      RMSE difference0.2330.0282.9021.8830.0200.002
    Tools

    Get Citation

    Copy Citation Text

    Zeyu ZHENG, Chao YE, Yucong ZHOU, Chunlian ZHAN, Han GAO. Calibration Method of Instrument Matrix for Polarization Metagrating[J]. Acta Photonica Sinica, 2025, 54(5): 0512002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Mar. 19, 2025

    Accepted: Apr. 17, 2025

    Published Online: Jun. 18, 2025

    The Author Email: Han GAO (gaohan@cjlu.edu.cn)