Spectroscopy and Spectral Analysis, Volume. 42, Issue 11, 3461(2022)

Traceable Measurement of Optical Path Length of Gas Cell Based on Tunable Diode Laser Absorption Spectroscopy

Jiang-xiong LONG1、*, Yu-jun ZHANG1、1; *;, Li SHAO1、1; *;, Qing YE1、1; 2;, Ying HE3、3;, Kun YOU3、3;, and Xiao-quan SUN1、1; 2;
Author Affiliations
  • 11. State Key Laboratory of Pulsed Power Laser Technology, National University of Defense Technology, Hefei 230037, China
  • 33. Hefei Institutes of Physics Science, Chinese Academy of Sciences, Hefei 230031, China
  • show less
    Figures & Tables(7)
    Schematic diagram of setup for optical path measurement of gas cellDFB: distributed feedback; D: detector (InGaAs); VG: Vacuum gauge; BV: Ball valve; NV: Needle valve; GC: Gas cylinder
    Output wavenumber of DFB as a function of temperature and current
    Detector signal and its baseline fitting at 74.2 torr and wavenumber corresponding to the sample point
    (a) Spectral absorbance and qSDVP fitting, (b) VP fitting residual, (c) qSDVP fitting residual at 74.2 Torr
    Integral absorbance A as a function of Γ and its general linear regression
    • Table 1. Integral absorbance and corresponding RSU at various pressure

      View table
      View in Article

      Table 1. Integral absorbance and corresponding RSU at various pressure

      p/TorrAfit/cm-1RSUfit/%RSUA/%
      36.0(6.527 8±0.023 1)×10-40.350.49
      48.7(8.835 7±0.028 8)×10-40.330.48
      59.4(1.076 4±0.003 4)×10-30.320.47
      74.2(1.343 6±0.003 7)×10-30.280.44
    • Table 2. Uncertainty budgets for optical path length L

      View table
      View in Article

      Table 2. Uncertainty budgets for optical path length L

      Varible (Unit)ValueRSU/%Contribution to the uncertainty of L/%
      Afit/cm-11.343 6×10-30.2829.10
      Kn10.3443.98
      S1,T0/(cm·molecule-1)6.647×10-240.158.38
      K1, T1.014 60.051.05
      S2,T0/(cm·molecule-1)9.768×10-2720.00
      K2, T0.973 60.050.00
      p/Pa9.892 5×1030.214.84
      riso1.0000.050.92
      χCO2/(mol·mol-1)10.0010.00
      T/K294.20.071.72
      kB/(J·K-1)1.380 648 53×10-2300
      L/cm81.61±0.420.51
    Tools

    Get Citation

    Copy Citation Text

    Jiang-xiong LONG, Yu-jun ZHANG, Li SHAO, Qing YE, Ying HE, Kun YOU, Xiao-quan SUN. Traceable Measurement of Optical Path Length of Gas Cell Based on Tunable Diode Laser Absorption Spectroscopy[J]. Spectroscopy and Spectral Analysis, 2022, 42(11): 3461

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Sep. 24, 2021

    Accepted: --

    Published Online: Nov. 23, 2022

    The Author Email: Jiang-xiong LONG (longjx@nudt.edu.cn)

    DOI:10.3964/j.issn.1000-0593(2022)11-3461-06

    Topics