Chinese Journal of Lasers, Volume. 46, Issue 10, 1004004(2019)

Research on Calibration System of Star Sensor Thermal Drift Based on Error Decoupling

Jingjing Gao1,2, Yong Bi1、*, Jinpeng Li1, Sen Pan3, and Xiaohan Yang3
Author Affiliations
  • 1Nanjing Research Centre of Astronomical Instruments, Chinese Academy of Sciences, Nanjing, Jiangsu 210042, China
  • 2University of Science and Technology of China, Hefei, Anhui 230026, China
  • 3Jiangsu University, Zhenjiang, Jiangsu 212013, China
  • show less
    Figures & Tables(8)
    Working principle diagram of thermal drift calibration system
    Optical design layout
    MTF of optical system
    Wavefront function curves of optical system. (a) 0°, 0°; (b) 0°, 0.8°; (c) 0°, 1.6°; (d) 0°, 4.5°; (e) 0°, 6°; (f) 0°, 8.5°
    Three measurement coordinate system of star sensor
    Simulation results of star sensor. (a) -25-20 ℃ mounting surface deformation; (b) -25-20 ℃ flange surface deformation; (c) 20-60 ℃ mounting surface deformation; (d) 20-60 ℃ flange surface deformation
    • Table 1. Design parameters for optical system

      View table

      Table 1. Design parameters for optical system

      ParameterContent
      Star map field of view≥12°×12°
      Spectral range400-1100 nm
      Exit distance≥400 mm
      Equipment length≤1000 mm
      Light source stability2%(one cycle)
      Operating temperature range10-30 ℃
    • Table 2. Deformation angle results of different temperature ranges

      View table

      Table 2. Deformation angle results of different temperature ranges

      TemperatureSurface deformationOffset around theX axis /[(″)·℃-1]Offset around theY axis /[(″)·℃-1]Offset around theZ axis /[(″)·℃-1]
      Mounting surface-0.072410325-0.047455354-0.055023665
      -25-20 ℃Flange surface-0.1847076940.000204549-0.221526503
      Ideal result-0.128865833-0.059728291-0.174889345
      Mounting surface0.0000818190.010431996-0.00572737
      20-60 ℃Flange surface0.026386813-0.1421615140.0462637159
      Ideal result0.025159520.1478888840.064864912
    Tools

    Get Citation

    Copy Citation Text

    Jingjing Gao, Yong Bi, Jinpeng Li, Sen Pan, Xiaohan Yang. Research on Calibration System of Star Sensor Thermal Drift Based on Error Decoupling[J]. Chinese Journal of Lasers, 2019, 46(10): 1004004

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: measurement and metrology

    Received: May. 13, 2019

    Accepted: Jun. 17, 2019

    Published Online: Oct. 25, 2019

    The Author Email: Bi Yong (yongb@nairc.ac.cn)

    DOI:10.3788/CJL201946.1004004

    Topics