Acta Optica Sinica, Volume. 39, Issue 9, 0904001(2019)

Radiation Calibration Method Using Uncooled Long-Wave Infrared Spectrometer

Yue Wu1,2, Zhizhen Zhu1,2, Yonghua Fang1,2、*, Leilei Zhang2, Wenkang Yang1,2, Mengqi Tao1,2, and Zhiqiang Ning2
Author Affiliations
  • 1 Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • 2 University of Science and Technology of China, Hefei, Anhui 230026, China
  • show less
    Figures & Tables(9)
    Uncooled long-wave infrared spectrometer. (a) Schematic; (b) structural diagram
    Schematic of radiation calibration using uncooled infrared spectrometer
    Photos of experimental setup for radiation calibration using uncooled infrared spectrometer
    Relations between radiances LS(Tb) of standard area-source blackbody and output gray values of system at different ambient temperatures
    Relations between radiances LS(Tb) of standard area-source blackbody and output gray values of system obtained by drift compensation at different ambient temperatures
    Errors between output gray values obtained by drift compensation at different ambient temperatures and theoretically calculated values at laboratory ambient temperature (25 ℃)
    • Table 1. Parameters of uncooled long-wave infrared spectrometer

      View table

      Table 1. Parameters of uncooled long-wave infrared spectrometer

      ParameterValue
      Spectrometer size /(mm×mm×mm)176×173×62
      Wavelength range /μm8-12
      Spectral resolution /nm80
      Numerical aperture0.22
      Detector array size64×1
    • Table 2. Parameters of standard area-source blackbody

      View table

      Table 2. Parameters of standard area-source blackbody

      ParameterValue
      Temperature range /℃0-100
      Emitter size /(mm×mm)102×102
      Emissivity0.97±0.02
      Stability /℃±0.003
      Temperature resolution /℃±0.03
    • Table 3. Correction results of output gray values of system at different ambient temperatures

      View table

      Table 3. Correction results of output gray values of system at different ambient temperatures

      Tb /℃D(T25)Te=20 ℃Te=30 ℃Te=35 ℃Te=40 ℃
      D(T20)D(T20,T25)ET20(T25)/%D(T30)D(T30,T25)ET30(T25)/%D(T35)D(T35,T25)ET35(T25)/%D(T40)D(T40,T25)ET40(T25)/%
      20256023772543.5-0.627422575.50.629192530.5-1.231072552-0.3
      302840.826052771.5-2.429712804.5-1.232032814.5-0.933372782-2.0
      403168.429903156.5-0.433453178.50.335233134.5-1.137003145-0.7
      503542.833113477.5-1.836873520.5-0.638813492.5-1.440633508-0.9
    Tools

    Get Citation

    Copy Citation Text

    Yue Wu, Zhizhen Zhu, Yonghua Fang, Leilei Zhang, Wenkang Yang, Mengqi Tao, Zhiqiang Ning. Radiation Calibration Method Using Uncooled Long-Wave Infrared Spectrometer[J]. Acta Optica Sinica, 2019, 39(9): 0904001