Laser & Optoelectronics Progress, Volume. 58, Issue 8, 0800002(2021)

Research Progress of Early Disease Detection Technology Based on Infrared Thermography

Mingzhu Zhao1, Yan Zhang1,2、*, and Yingyan Zhu1
Author Affiliations
  • 1College of Big Data and Information Engineering, Guizhou University, Guiyang,Guizhou 550025, China
  • 2Research Center of Nondestructive Testing for Agricultural Products, Guiyang University, Guiyang, Guizhou 550005,China
  • show less
    Figures & Tables(10)
    Workflow of infrared thermal imaging detection system
    Infrared thermograms of breast masses[6]. (a)Patient with invasive ductal carcinoma; (b)patient with fibroadenoma; (c)patient with cyst
    Skin cancer detection algorithm block diagram
    Comparison of infrared thermograms of hands in patients with carpal tunnel syndrome[12].(a)Before operation; (b)after operation;(c)healthy volunteer
    Process of myocardial ischemic pre-diagnosis based on infrared imaging data[21]
    Infrared thermograms of face.(a)Patient with left-sided peripheral facial paralysis;(b)patient with central facial paralysis;(c)healthy volunteer
    Basic structure diagram of convolutional neural network
    • Table 1. Comparison of mammography, ultrasonography, and far-infrared thermography in the diagnosis of breast lesions less than 2 cm in diameter (n=853)[18]
      View table

      Table 1. Comparison of mammography, ultrasonography, and far-infrared thermography in the diagnosis of breast lesions less than 2 cm in diameter (n=853)[18]
      ItemPathological examinationAccuracy/%Sensitivity/%Specificity/%
      MalignantBenign
      Mammography
      Malignant591996.180.897.6
      Benign14761
      Color Doppler ultrasonography
      Malignant632695.886.396.6
      Benign10754
      Far-infrared thermography
      Malignant661797.190.497.8
      Benign7763

    • Table 2. Comparison of mammography, ultrasonography, and far-infrared thermography in the diagnosis of breast lesions large than 2 cm in diameter (n=1193)[18]

      View table

      Table 2. Comparison of mammography, ultrasonography, and far-infrared thermography in the diagnosis of breast lesions large than 2 cm in diameter (n=1193)[18]

      ItemPathological examinationAccuracy/%Sensitivity/%Specificity/%
      MalignantBenign
      Mammography
      Malignant35411685.084.985.1
      Benign63660
      Color Doppler ultrasonography
      Malignant3967891.795.090.0
      Benign21698
      Far-infrared thermography
      Malignant3753993.290.095.0
      Benign42737
    • Table 3. Comparison of breast cancer classification results based on differential thermal images

      View table

      Table 3. Comparison of breast cancer classification results based on differential thermal images

      ItemAccuracy/%Sensitivity/%Specificity/%
      Original images
      CNN62.823.9100
      Differential thermal images
      CNN73.497.865.2
      AdaBoost87.291.383.3
    Tools

    Get Citation

    Copy Citation Text

    Mingzhu Zhao, Yan Zhang, Yingyan Zhu. Research Progress of Early Disease Detection Technology Based on Infrared Thermography[J]. Laser & Optoelectronics Progress, 2021, 58(8): 0800002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Reviews

    Received: Aug. 17, 2020

    Accepted: Sep. 15, 2020

    Published Online: Apr. 12, 2021

    The Author Email: Yan Zhang (Eileen_zy001@sohu.com)

    DOI:10.3788/LOP202158.0800002

    Topics