Chinese Journal of Lasers, Volume. 44, Issue 11, 1111002(2017)

Quantitative Detection of Nutrient Elements in Soil Based on Laser Induced Breakdown Spectroscopy

Wang Jinmei, Yan Haiying, Zheng Peichao*, and Tan Guining
Author Affiliations
  • [in Chinese]
  • show less
    Figures & Tables(10)
    (a) Traditional LIBS system and (b) LIBS system with cage structure
    Variation curves of signal intensity of Mg II 279.56 nm, Mg II 280.26 nm, and Mg I 285.21 nmspectral lines at different focusing positions
    Plasma images at different focusing positions
    Stability of Mg II 279.56 nm, Mg II 280.26 nm, and Mg I 285.21 nm spectral lines in traditional LIBS system
    Stability of Mg II 279.56 nm, Mg II 280.26 nm, and Mg I 285.21 nm spectral lines in LIBS system with cage structure
    Calibration curves for different elements. (a) Copper element; (b) manganese element; (c) magnesium element; (d) potassium element
    • Table 1. Mass fractions of Cu, Mn, Mg and K elements in soil samples10-6

      View table

      Table 1. Mass fractions of Cu, Mn, Mg and K elements in soil samples10-6

      NumberCuMnMgKNumberCuMnMgK
      124.6677.015300.017600.0944.2863.016100.018933.3
      254.0956.016500.019600.01022.5703.013140.019950.0
      318.3755.08820.024650.01120.4729.010980.022300.0
      439.3816.515900.018600.01242.1889.013940.021283.3
      521.5716.012060.021125.01334.4770.015700.018266.7
      636.2855.512660.022125.01430.4766.313980.019862.5
      732.3796.013540.020616.71537.7836.014280.020362.5
      830.2822.011380.022966.71628.8785.812360.021625.0
    • Table 2. Actual laser pulse energy for 20 timesmJ

      View table

      Table 2. Actual laser pulse energy for 20 timesmJ

      NumberEnergyNumberEnergyNumberEnergyNumberEnergyNumberEnergy
      150.5551.9951.11350.61749.6
      250.8648.91050.61451.51850.5
      351.6749.61151.01550.01948.9
      449.6851.31252.01650.22049.6
    • Table 3. Detection limits of Cu, Mn, Mg and K elements obtained by LIBS system with cage structure and traditional LIBS system(mass fraction)10-6

      View table

      Table 3. Detection limits of Cu, Mn, Mg and K elements obtained by LIBS system with cage structure and traditional LIBS system(mass fraction)10-6

      ElementCuMnMgK
      LIBS system with cage structure0.4213.238.562
      Traditional LIBS system10.58145.2556.280.56
    • Table 4. Predicted mass fractions and relative errors of Cu, Mn, Mg, K elements in soil samples

      View table

      Table 4. Predicted mass fractions and relative errors of Cu, Mn, Mg, K elements in soil samples

      NumberPredicted mass fraction /10-6Relative error /%
      CuMnMgKCuMnMgK
      948.11021.317363.517170.88.115.57.310.3
      1022.9623.714219.021741.61.612.77.68.2
      1121.5737.212265.425488.55.21.110.512.5
      1248.01005.215681.023653.912.311.611.110.0
      1339.2891.116531.020979.212.313.65.012.9
      1433.6694.015385.921978.39.710.49.19.6
      1542.0838.015934.722784.510.20.210.410.6
      1633.5705.13246.524883.013.911.46.713.1
    Tools

    Get Citation

    Copy Citation Text

    Wang Jinmei, Yan Haiying, Zheng Peichao, Tan Guining. Quantitative Detection of Nutrient Elements in Soil Based on Laser Induced Breakdown Spectroscopy[J]. Chinese Journal of Lasers, 2017, 44(11): 1111002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Spectroscopy

    Received: Jun. 20, 2017

    Accepted: --

    Published Online: Nov. 17, 2017

    The Author Email: Peichao Zheng (Zhengpc@cqupt.edu.cn)

    DOI:10.3788/CJL201744.1111002

    Topics