Chinese Journal of Lasers, Volume. 44, Issue 10, 1004003(2017)

Extrinsic Parameters Calibration of Three-Dimensional Measurement System for Ultra-Large Scale Line-Structured Light Sensor

Xie Zexiao*, Zhu Ruixin, and Zhang Anqi
Author Affiliations
  • [in Chinese]
  • show less
    Figures & Tables(13)
    Three-dimensional measurement system with ultra-large scale line-structured light sensor.(a) Measurement range of line-structured light sensor; (b) physical map of system
    Mathematical model of line-structured light scanning measurement system
    Physical map of one-dimensional target
    Schematic of extrinsic parameter calibration principle in line-structured light sensor
    Three standard balls for accuracy test of system
    Point cloud data on standard ball
    Error distribution between fitted sphere and point cloud
    Error analysis of system. (a) Standard deviation of point clouds; (b) ball radius error
    Physical map of measured work piece
    Measurement effect image of measured work piece
    • Table 1. Extrinsic-parameter calibration results of line-structured light sensor

      View table

      Table 1. Extrinsic-parameter calibration results of line-structured light sensor

      Calibration No.Xm-axis calibration resultYm-axis calibration result
      cosθxcosθycosθzcosθxcosθycosθz
      10.003299-0.8418630.539682-0.999865-0.0138070.008895
      20.003291-0.8417900.539796-0.999874-0.0133570.008542
      30.003229-0.8418630.539682-0.999883-0.0130620.007935
    • Table 2. Maximum deviations between fitted sphere and point cloud in positive and negative normal directions

      View table

      Table 2. Maximum deviations between fitted sphere and point cloud in positive and negative normal directions

      No. of ball positionBall nameProposed methodCalibration based on planar target
      d1 /mmd2 /mmd1 /mmd2 /mm
      1A1-0.23050.2469-0.23610.2442
      2B1-0.25050.2147-0.24740.2533
      3C1-0.23300.2343-0.26350.2546
      4A2-0.26670.2525-0.28250.2502
      5B2-0.27210.2550-0.27240.2495
      6C2-0.26920.2554-0.24270.2451
      7A3-0.24690.2346-0.24740.2327
      8B3-0.21000.2450-0.20280.2444
      9C3-0.24090.2108-0.23430.2207
      10A4-0.25060.2268-0.26330.2370
      11B4-0.23850.2379-0.27380.2547
      12C4-0.23840.2392-0.26210.2505
      13A5-0.26370.2503-0.26490.2497
      14B5-0.27720.2578-0.26910.2801
      15C5-0.26410.2516-0.24800.2481
      Maximum deviation-0.27720.2578-0.28250.2801
    • Table 3. Errors of sphere centre distance

      View table

      Table 3. Errors of sphere centre distance

      Name of ballsNo. of ballpositionProposed methodCalibration based on planar target
      M/mmE/mmM/mmE/mm
      1152.36320.3002152.37280.3098
      2152.31530.2523152.34030.2773
      AB3152.32820.2652152.32950.2665
      4152.26020.1972152.30610.2431
      5152.25460.1916152.26480.2018
      1220.92600.5470220.88530.5063
      2220.82600.4470220.78250.4035
      AC3221.00320.6242220.98110.6021
      4220.80630.4273220.75130.3723
      5220.70080.3218220.68960.3106
      1222.20100.4110222.21730.4273
      2222.08180.2918222.07790.2879
      BC3222.34900.5590222.36120.5712
      4222.09620.3062222.08600.2960
      5221.93010.1401221.99810.2081
      Maximum deviation-0.6242-0.6021
    Tools

    Get Citation

    Copy Citation Text

    Xie Zexiao, Zhu Ruixin, Zhang Anqi. Extrinsic Parameters Calibration of Three-Dimensional Measurement System for Ultra-Large Scale Line-Structured Light Sensor[J]. Chinese Journal of Lasers, 2017, 44(10): 1004003

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: measurement and metrology

    Received: May. 4, 2017

    Accepted: --

    Published Online: Oct. 18, 2017

    The Author Email: Zexiao Xie (xiezexiao@ouc.edu.cn)

    DOI:10.3788/CJL201744.1004003

    Topics