Acta Optica Sinica, Volume. 44, Issue 11, 1130002(2024)

Improvement of Detection Stability of Laser-Induced Breakdown Spectroscopy Based on Two-Dimensional Wavelet Denoising

Ziyi Zhao1,2, Zhongqi Hao1,2、*, Ying Lu1,2, Zhishuai Xu1,2, Baining Xu1,2, Neng Zhang1,2, Li Liu1,2, Jiulin Shi1,2、**, and Xingdao He1,2
Author Affiliations
  • 1Key Laboratory of Opto-Electronic Information Science and Technology of Jiangxi Province, Nanchang Hangkong University, Nanchang 330063, Jiangxi , China
  • 2Key Laboratory of Non-Destructive Testing of Ministry of Education, Nanchang Hangkong University, Nanchang 330063, Jiangxi , China
  • show less
    Figures & Tables(9)
    Schematic of LIBS setup
    Multiwavelet decomposition process. (a) One-dimensional wavelet; (b) two-dimensional wavelet
    SNR and RSD of element characteristic lines versus layer number. (a) SNR; (b) RSD
    Comparison of spectra before and after two wavelet denoising. (a) Raw data; (b) data processed by one-dimensional wavelet; (c) data processed by two-dimensional wavelet
    Comparison of SNR of spectral data after wavelet denoising with original data for different elements. (a) Cu; (b) Ni; (c) Mo; (d) V
    Comparison of RSD of spectral data after wavelet denoising with the original data for different elements. (a) Cu; (b) Ni; (c) Mo; (d) V
    Quantitative analysis results of four elements. (a) Cu; (b) Ni; (c) Mo; (d) V
    • Table 1. Mass fraction of Cu, Ni, Mo, and V in the standard sample of alloy structural steel

      View table

      Table 1. Mass fraction of Cu, Ni, Mo, and V in the standard sample of alloy structural steel

      Sample numberMass fraction /%
      NiCuMoV
      GBW 01666a3.5300.3550.2100.189
      GBW 01667a4.3300.0160.0190.478
      GBW 01668a1.2900.1360.5560.346
      GBW 01669a2.5400.4411.4000.449
      GBW 01670a1.7700.0460.8640.118
      GBW 01671a1.0200.2450.1180.273
      GBW 01672a0.7730.5180.3050.013
      GBW 01673a0.6210.6360.4230.055
    • Table 2. Comparison of the coefficient of determination of calibration curves before and after the denoising of two wavelets

      View table

      Table 2. Comparison of the coefficient of determination of calibration curves before and after the denoising of two wavelets

      Processing methodCoefficient of determination
      CuNiMoV
      Raw data0.96480.95730.91560.9936
      One-dimensional wavelet0.94540.95710.92210.9934
      Two-dimensional wavelet0.98110.95860.92900.9944
    Tools

    Get Citation

    Copy Citation Text

    Ziyi Zhao, Zhongqi Hao, Ying Lu, Zhishuai Xu, Baining Xu, Neng Zhang, Li Liu, Jiulin Shi, Xingdao He. Improvement of Detection Stability of Laser-Induced Breakdown Spectroscopy Based on Two-Dimensional Wavelet Denoising[J]. Acta Optica Sinica, 2024, 44(11): 1130002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Spectroscopy

    Received: Dec. 25, 2023

    Accepted: Mar. 12, 2024

    Published Online: Jun. 17, 2024

    The Author Email: Hao Zhongqi (hzq@nchu.edu.cn), Shi Jiulin (jlshi@nchu.edu.cn)