Acta Optica Sinica, Volume. 42, Issue 5, 0512006(2022)

Experimental Study on Traceability of Absolute Radiometer to Cryogenic Radiometer

Junchao Zhang*, Nan Xu, Ning Xu, Haifeng Meng, Wende Liu, Chuan Cai, Bifeng Zhang, Shuai Man, Meng Wang, Yinliang Li, and Limin Xiong
Author Affiliations
  • Division of Metrology in Optics and Laser, National Institute of Metrology, Beijing 100029, China
  • show less
    Figures & Tables(6)
    Reflectivity curve of black coating in 250--2500 nm
    Experimental device diagram of trap detector tracing to cryogenic radiometer
    Experimental device diagram of absolute radiometer tracing to trap detector
    • Table 1. Calibration data of absolute power responsivity of trap detector (0901#)

      View table

      Table 1. Calibration data of absolute power responsivity of trap detector (0901#)

      Wavelength /nmPower responsivity /( A·W-1)
      4880.390860
      5140.412614
      6330.508590
      6470.520140
      7520.604980
      7990.642560
    • Table 2. Irradiance output at different positions of absolute radiometer

      View table

      Table 2. Irradiance output at different positions of absolute radiometer

      Spot positionIrradiance /(W·m-2)
      Single measurement valueAverage value
      Center position977.9978.7977.9978.2
      Move to left (1 mm)981.0980.7980.2980.6
      Move to right (1 mm)975.5975.1975.6975.4
      Move up (1 mm)982.3982.4982.6982.4
      Move down (1 mm)980.8981.2981.7981.2
    • Table 3. Uncertainty assess of irradiance calibration of absolute radiometer

      View table

      Table 3. Uncertainty assess of irradiance calibration of absolute radiometer

      UncertaintycomponentSource of uncertaintyUncertaintytypeUncertaintydistributionRelativeuncertainty /%
      u1Measurement repeatabilityANormal0.030
      u2Light source instabilityBEquidistribution0.002
      u3Trap detector calibrationBNormal0.015
      u4Interpolation calculation of standard value of trap detectorBEquidistribution0.030
      u5Nonlinearity of signal amplifier and monitoring detectorBEquidistribution0.050
      u6Absorptivity and wavelength flatness of black coating ofabsolute radiometerBEquidistribution0.030
      u7Absolute radiometer position response non-uniformityBEquidistribution0.410
      u8Aperture calibrationBNormal0.100
    Tools

    Get Citation

    Copy Citation Text

    Junchao Zhang, Nan Xu, Ning Xu, Haifeng Meng, Wende Liu, Chuan Cai, Bifeng Zhang, Shuai Man, Meng Wang, Yinliang Li, Limin Xiong. Experimental Study on Traceability of Absolute Radiometer to Cryogenic Radiometer[J]. Acta Optica Sinica, 2022, 42(5): 0512006

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Aug. 17, 2021

    Accepted: Sep. 26, 2021

    Published Online: Mar. 8, 2022

    The Author Email: Zhang Junchao (zhangjc@nim.ac.cn)

    DOI:10.3788/AOS202242.0512006

    Topics