NUCLEAR TECHNIQUES, Volume. 47, Issue 12, 120403(2024)

Uncertainty analysis method and application of neutron multi-sphere spectrometer measurement based on Monte Carlo method

Dake ZHANG1,2,3, Zhihui TANG1,2,3、*, Yingjing WEI1,2,3, Taosheng LI4, Qinjian CAO1, and Bowen SUN1,2,3
Author Affiliations
  • 1Institute of Health Physics, China Institute for Radiation Protection, Taiyuan 030006, China
  • 2Shanxi Provincial Key Laboratory of Radiation Safety and Protection, Taiyuan 030006, China
  • 3Key Laboratory of Radiation Protection Technology of China National Nuclear Corporation, Taiyuan 030006, China
  • 4Key Laboratory of Neutron Transport Theory and Radiation Safety, Institute of Nuclear Energy Safety Technology, Chinese Academy of Sciences, Hefei 230031, China
  • show less
    Figures & Tables(10)
    MC method assessment uncertainty flow chart
    SP94 thermal neutron detector (a) Snapshot of physical detector, (b) Blueprint
    Photo of irradiation measurement site
    Distribution of fitting results
    Spectrum unfolding results with error bars
    • Table 1. Data related to the neutron source used to reference the radiation

      View table
      View in Article

      Table 1. Data related to the neutron source used to reference the radiation

      源的类型

      Type of source

      活度

      Activity / Bq

      注量平均能量

      Average energy fluence / MeV

      半衰期

      Half-life period / a

      241Am-Be3.7×10114.16432
      241Am-Be3.7×1094.16432
      252Cf3.7×1092.132.65
      252Cf3.7×1072.132.65
      252Cf+D2O3.7×1090.552.65
    • Table 2. Relative spread uncertainty of neutron sources for reference radiation

      View table
      View in Article

      Table 2. Relative spread uncertainty of neutron sources for reference radiation

      源的类型

      Type of source

      注量平均能量

      Average energy fluence / MeV

      相对扩展不确定度

      Reference expended uncertainty Urel (k=2) / %

      241Am-Be4.164.6
      241Am-Be4.167.0
      252Cf2.135.0
      252Cf2.135.0
      252Cf+D2O0.555.0
    • Table 3. Fitting results

      View table
      View in Article

      Table 3. Fitting results

      拟合结果

      Fitting result

      正态

      Normal

      均值Mean value98
      方差Dispersion3.78
    • Table 4. Fitting result parameters

      View table
      View in Article

      Table 4. Fitting result parameters

      参数

      Parameter

      估计值

      Estimated value

      标准误差

      Standard deviation

      μ9.80×1011.94×10-2
      σ1.44×10-181.89×10-4
    • Table 5. Estimated covariances for parameter estimates

      View table
      View in Article

      Table 5. Estimated covariances for parameter estimates

      关系

      Relationship

      协方差

      Covariance

      μ-μ3.78×10-4
      μ-σ1.44×10-18
      σ-σ1.89×10-4
    Tools

    Get Citation

    Copy Citation Text

    Dake ZHANG, Zhihui TANG, Yingjing WEI, Taosheng LI, Qinjian CAO, Bowen SUN. Uncertainty analysis method and application of neutron multi-sphere spectrometer measurement based on Monte Carlo method[J]. NUCLEAR TECHNIQUES, 2024, 47(12): 120403

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: NUCLEAR ELECTRONICS AND INSTRUMENTATION

    Received: Mar. 28, 2024

    Accepted: --

    Published Online: Jan. 15, 2025

    The Author Email: TANG Zhihui (TANGZhihui)

    DOI:10.11889/j.0253-3219.2024.hjs.47.120403

    Topics