Laser & Optoelectronics Progress, Volume. 61, Issue 13, 1317001(2024)

Wavelength Optimization and Experimental Validation of Near-Infrared Tissue Oxygenation Detection

Xiang Shi1, Yao Huang2,3、*, Jingshu Ni2, Meili Dong2,3, Yang Zhang2,3, Yuanzhi Zhang2,3, Yikun Wang1,2,3、**, and Yong Liu2,3
Author Affiliations
  • 1Department of Biomedical Engineering, Anhui Medical University, Hefei 230009, Anhui, China
  • 2Anhui Medical Optical Diagnosis and Treatment Technology and Equipment Engineering Laboratory, Anhui Biomedical Optical Instrument Engineering Technology Research Center, Anhui Institute of Optics and Fine Mechanics, Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei 230031, Anhui, China
  • 3Wanjiang Center for Development of Emerging Industrial Technology, Tongling 244000, Anhui, China
  • show less
    Figures & Tables(14)
    Transmission system
    Scattering system
    Absorption coefficient of hemoglobin
    Schematic diagram of tissue blood oxygen detection
    Statistical results of condition number
    The results of fat emulsion phantom experiments
    Experimental results of melanin phantom.
    The results of noise reduction. (a) The original figure; (b) smoothing filtering; (c) wavelet transform; (d) Fourier transform method
    The results of the error contrast. (a) Smoothing filtering; (b) wavelet transform; (c) Fourier transform method
    • Table 1. The mass concentration ratio of blood phantom

      View table

      Table 1. The mass concentration ratio of blood phantom

      Experimental sequenceFat emulsion /%Hemoglobin /(mg/mL)
      12.55
      25.05
      37.55
      410.05
    • Table 2. The mass concentration ratio of melanin phantom

      View table

      Table 2. The mass concentration ratio of melanin phantom

      Experimental sequenceFat emulsion /%Hemoglobin /(mg/mL)Melanin /(mg/L)
      12.5525
      22.5550
      32.5575
      42.55100
      52.55125
      62.55150
    • Table 3. The results of error analysis

      View table

      Table 3. The results of error analysis

      No.AVG /(mg/mL)MAE /(mg/mL)MSE /(mg2/mL2
      Set15.500.620.90
      Set26.061.503.21
      Set36.631.784.34
    • Table 4. The results of error analysis

      View table

      Table 4. The results of error analysis

      No.AVG /(mg/mL)MAE /(mg/mL)MSE /(mg2/mL2
      Set14.960.190.08
      Set26.761.765.74
      Set36.762.0111.11
    • Table 5. Absolute mean error for the combinations of the three wavelengths

      View table

      Table 5. Absolute mean error for the combinations of the three wavelengths

      No.Smoothing filteringWavelet transformFourier transform method
      Set10.331.981.70
      Set20.322.521.79
      Set322.6636.4948.57
    Tools

    Get Citation

    Copy Citation Text

    Xiang Shi, Yao Huang, Jingshu Ni, Meili Dong, Yang Zhang, Yuanzhi Zhang, Yikun Wang, Yong Liu. Wavelength Optimization and Experimental Validation of Near-Infrared Tissue Oxygenation Detection[J]. Laser & Optoelectronics Progress, 2024, 61(13): 1317001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Medical Optics and Biotechnology

    Received: Jan. 31, 2024

    Accepted: Mar. 5, 2024

    Published Online: Jul. 4, 2024

    The Author Email: Yao Huang (yhuang@aiofm.ac.cn), Yikun Wang (wyk@aiofm.ac.cn)

    DOI:10.3788/LOP240650

    CSTR:32186.14.LOP240650

    Topics