Chinese Journal of Lasers, Volume. 45, Issue 8, 804005(2018)

Quantum Efficiency Calibration of UV-VUV Photomultiplier Tube

Li Hanshuang1,2, Li Bo1, and Wang Shurong1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less
    Figures & Tables(6)
    Electrical connection diagram of phototube
    Light path diagram of vacuum system
    Quantum efficiency curves of standard phototube and phototube R2078
    • Table 1. Data of standard vacuum phototube

      View table

      Table 1. Data of standard vacuum phototube

      λ /nmIk /(10-10 A)Lp /(10-10 W)η /%
      1502.42511.887210.643
      1703.53351.736014.875
      1906.48352.254418.805
      2107.98542.499118.904
      2308.24382.420618.396
      2506.46421.980016.225
    • Table 2. Data of phototube R2078

      View table

      Table 2. Data of phototube R2078

      λ /nmIk /(10-10 A)Lp /(10-10 W)η /%
      1500.18781.88720.824
      1701.82871.73607.698
      1903.67792.254410.668
      2104.82372.499111.419
      2305.69162.420612.701
      2504.69181.980011.776
    • Table 3. Estimated calibration uncertainty of phototube R2078

      View table

      Table 3. Estimated calibration uncertainty of phototube R2078

      Source of uncertainyUncertainty estimation /%
      Deuterium stabilization2.5
      Monochromator repeatability0.5
      Uncertainty of standard phototube2
      Uncertainty of phototube R20781
      Association uncertainty3.4
    Tools

    Get Citation

    Copy Citation Text

    Li Hanshuang, Li Bo, Wang Shurong. Quantum Efficiency Calibration of UV-VUV Photomultiplier Tube[J]. Chinese Journal of Lasers, 2018, 45(8): 804005

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Measurement and metrology

    Received: Jan. 19, 2018

    Accepted: --

    Published Online: Aug. 11, 2018

    The Author Email:

    DOI:10.3788/CJL201845.0804005

    Topics