NUCLEAR TECHNIQUES, Volume. 45, Issue 12, 120401(2022)

Highly sensitive coded aperture gamma imaging system based on a SPECT probe

Xinyao LIU1, Qingyang WEI1、*, Tianpeng XU2, and Zhaohui ZHANG1
Author Affiliations
  • 1Beijing Engineering Research Center of Industrial Spectrum Imaging, School of Automation and Electrical Engineering, University of Science and Technology Beijing, Beijing 100083, China
  • 2State Nuclear Security Technology Center, Beijing 102445, China
  • show less
    Figures & Tables(10)
    Flow chart of physical detection process simulation of coded aperture γ imaging system
    Pattern of the coded mask: MURA(45×45)
    Illustration of the geometry setup of the imaging system
    Two-dimensional model of photon counting estimation for detection
    Reconstructed images of a point source at different distances and different FOV angles: the 1st, 2nd, 3rd, 4th, 5th and 6th columns are positions at (-29.64°, 23.65°), (0°, 23.65°), (0°, 0°), (28.66°, 0°), (29.64°, 0°) and (-14.87°, -14.87°); the 1st, 2nd and 3rd rows are the distance between the radiation source and the detector of 10 m, 15 m and 18 m, respectively
    Reconstruction images of 1.11×106 Bq 137Cs at distance of 4 m from detector with different FOV angles: the 1st, 2nd, 3rd, 4th and 5th columns are positions at (-29.64°, 23.65°), (0°, 23.65°), (0°, 0°), (28.66°, 0°) and (-14.87°, -14.87°), respectively
    (a) Reconstructed image of 3.7×107 Bq 137Cs point source at distance of 10 m from detector with FOV angle (0°, 0°), Gaussian fitting of normalized data points of pixel intensity distribution along theX direction (b) and theY direction (c) of a single point source
    Reconstructed images of two 3.7×107 Bq 137Cs sources at the distance of 10 m from detector with acquisition time of 1 s
    • Table 1. Parameters of imaging system

      View table
      View in Article

      Table 1. Parameters of imaging system

      参数Parameter优化设计值Optimal design value
      探测器Detector材料MaterialNaI晶体 NaI crystal
      单位像素大小Unit pixel size3.55 mm

      厚度Thickness

      阵列Array

      整体面积Overall area

      9.5 mm

      143×110

      510 mm×390 mm

      编码板Coded mask材料Material钨合金Tungsten alloy
      单位孔径Unit hole size20.1 mm×17 mm

      厚度Thickness

      MURA阵列 MURA array

      整体面积Overall area

      10 mm

      45×45

      1 020 mm×780 mm

      成像系统Imaging system焦距Focal length441 mm
    • Table 2. CNR of reconstructed images from different positions of radioactive sources

      View table
      View in Article

      Table 2. CNR of reconstructed images from different positions of radioactive sources

      放射源与成像系统的距离

      Distance of sources and the system / m

      放射源的视场角度FOV angle position of radiation source
      (-29.64°, 23.65°)(0°, 23.65°)(0°, 0°)(28.66°, 0°)(29.64°, 0°)(-14.87°, -14.87°)
      10107.43441.90207.83415.96136.85211.23
      1572.1632.9222.2335.3124.0747.46
      1819.3720.7813.5627.1630.3412.25
    Tools

    Get Citation

    Copy Citation Text

    Xinyao LIU, Qingyang WEI, Tianpeng XU, Zhaohui ZHANG. Highly sensitive coded aperture gamma imaging system based on a SPECT probe[J]. NUCLEAR TECHNIQUES, 2022, 45(12): 120401

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Jun. 18, 2022

    Accepted: --

    Published Online: Jan. 3, 2023

    The Author Email: WEI Qingyang (weiqy@ustb.edu.cn)

    DOI:10.11889/j.0253-3219.2022.hjs.45.120401