Chinese Journal of Lasers, Volume. 48, Issue 20, 2007001(2021)

Spatially Resolved Reflectance in Sub-Diffusive Regime Measurement System for Mucosal Lesion Recognition

Qi Zhang1, Limin Zhang1,2、*, Jiayu Wang1, Jingyi An1, Jing Wang3, chen Chen3, and Feng Gao1,2
Author Affiliations
  • 1College of Precision Instrument and Optoelectronics Engineering, Tianjin University, Tianjin 300072, China
  • 2Tianjin Key Laboratory of Biomedical Detecting Techniques and Instruments, Tianjin 300072, China
  • 3The First Affiliated Hospital of Xinjiang Medical University, Urumqi, Xinjiang 830054, China
  • show less
    Figures & Tables(9)
    Schematic of experimental system
    System stability measurement results
    Experimental multi-frequency crosstalk ratio
    Intensity amplitude measured at different ambient light intensities. (a) Without digital lock-in; (b) digital lock-in
    Linearity assessments of the system at source-detector separation of 220 μm
    Optical parameter inversion results. (a) μa; (b) μ's; (c)γ
    • Table 1. Correlation coefficient of linear fitting at four wavelengths

      View table

      Table 1. Correlation coefficient of linear fitting at four wavelengths

      Wavelength /nmSource-dectetor separation /μm
      2204406608801100
      5200.99920.99820.99770.99870.9984
      6500.99910.99940.99940.99920.9992
      7850.99880.99890.99910.99730.9985
      8300.99850.99930.99850.99880.9983
    • Table 2. Optical parameters of turbid phantoms

      View table

      Table 2. Optical parameters of turbid phantoms

      No.μa /cm-1μ's /cm-1γNo.μa /cm-1μ's /cm-1γ
      P10.00192.2507P70.5132.0110
      P20.001132.0110P80.1171.9992
      P30.001171.9992P90.5171.9992
      P40.001292.0110P101.5171.9992
      P50.192.2507P110.1292.0110
      P62.092.2507P122.0292.0110
    • Table 3. Root-mean-square error and relative root-mean-square error of optical parameters

      View table

      Table 3. Root-mean-square error and relative root-mean-square error of optical parameters

      ParameterRoot-mean-square errorRelative root-mean-square error /%
      μa0.102326.0074
      μ's0.54422.9835
      γ0.03931.9139
    Tools

    Get Citation

    Copy Citation Text

    Qi Zhang, Limin Zhang, Jiayu Wang, Jingyi An, Jing Wang, chen Chen, Feng Gao. Spatially Resolved Reflectance in Sub-Diffusive Regime Measurement System for Mucosal Lesion Recognition[J]. Chinese Journal of Lasers, 2021, 48(20): 2007001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: biomedical photonics and laser medicine

    Received: Jun. 16, 2021

    Accepted: Aug. 15, 2021

    Published Online: Oct. 12, 2021

    The Author Email: Limin Zhang (zhanglm@tju.edu.cn)

    DOI:10.3788/CJL202148.2007001

    Topics