NUCLEAR TECHNIQUES, Volume. 48, Issue 5, 050301(2025)

Study on background count rate and counting efficiency of plastic scintillation microspheres

Hui ZHANG1,2、*, Yadong WANG1,2, Yan MA1,2, and Lina MA1,2
Author Affiliations
  • 1China Institute for Radiation Protection, Taiyuan 030006, China
  • 2Shanxi Provincial Key Laboratory for Translational Nuclear Medicine and Precision Protection, Taiyuan 030006, China
  • show less
    Figures & Tables(8)
    Changes in background counting rate of PSm with shading time
    Variations of background count rate and 14C counting efficiency with the height of PSm sample
    Measured and expected values for PSm measurement of 14C
    • Table 1. Parameters of different liquid scintillation vials

      View table
      View in Article

      Table 1. Parameters of different liquid scintillation vials

      序号

      Number

      材质

      Material

      规格

      Specifications / mL

      外径

      Outer diameter

      / mm

      高度

      Height

      / mm

      G7玻璃Glass716.857.5
      G20玻璃Glass2027.260.7
      P7塑料Plastics714.657.5
      P20塑料Plastics2026.260.7
    • Table 2. Background counting rate and 14C counting efficiency of four types of liquid scintillation vials

      View table
      View in Article

      Table 2. Background counting rate and 14C counting efficiency of four types of liquid scintillation vials

      序号 Number背景计数率 Background count / Counts∙min-114C计数效率 Count efficiency of 14C / %
      G717.141.0
      G2022.639.5
      P717.341.8
      P2020.839.5
    • Table 3. Background counting rate and 14C counting efficiency for different PSm dosages

      View table
      View in Article

      Table 3. Background counting rate and 14C counting efficiency for different PSm dosages

      PSm质量

      Mass of PSm / g

      高度

      Height of PSm / mm

      背景计数

      Background count / Counts∙min-1

      14C计数效率

      Count efficiency of 14C / %

      2815.532.9
      41417.636.0
      62019.137.1
      82619.238.4
      103220.839.5
      123821.940.3
    • Table 4. MDA of different liquid scintillation vials

      View table
      View in Article

      Table 4. MDA of different liquid scintillation vials

      液闪瓶 Liquid scintillation vialsMDA / Bq·L-1
      G7131
      G2051
      P7133
      P2048
    • Table 5. Measurement of 14C specific activity using PSm

      View table
      View in Article

      Table 5. Measurement of 14C specific activity using PSm

      期望值

      Expected value

      / Bq·L-1

      测量值

      Measurement value

      / Bq·L-1

      偏差

      Deviation

      / %

      574.5±8.6545.9±43.7-5.0
      1 038.4±15.61 028.1±49.2-1.0
      2 284.6±34.32 331.8±69.52.1
      2 680.5±40.22 705.4±74.40.9
      3 825.6±57.43 853.9±86.20.7
    Tools

    Get Citation

    Copy Citation Text

    Hui ZHANG, Yadong WANG, Yan MA, Lina MA. Study on background count rate and counting efficiency of plastic scintillation microspheres[J]. NUCLEAR TECHNIQUES, 2025, 48(5): 050301

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: NUCLEAR CHEMISTRY, RADIOCHEMISTRY, RADIOPHARMACEUTICALS AND NUCLEAR MEDICINE

    Received: Jun. 17, 2024

    Accepted: --

    Published Online: Jun. 26, 2025

    The Author Email: Hui ZHANG (513563244@qq.com)

    DOI:10.11889/j.0253-3219.2025.hjs.48.240277

    Topics