Acta Optica Sinica, Volume. 44, Issue 7, 0728001(2024)

Dual-Channel Reflective Optical Microfiber Coupler Diaphragms for Continuous and Accurate Blood Pressure Monitoring

Xue Zou1,2, Junhao Fan2, Binbin Luo2、*, Fumin Zhou2, Decao Wu2, Zufan Zhang1, and Mingfu Zhao2
Author Affiliations
  • 1School of Communication and Information Engineering, Chongqing University of Posts and Telecommunications, Chongqing 400065, China
  • 2Chongqing Key Laboratory of Optical Fiber Sensor and Photoelectric Detection, Chongqing University of Technology, Chongqing 400054, China
  • show less
    Figures & Tables(14)
    Diagram of structure. (a) Structure of R-OMC; (b) R-OMC under microscope; (c) structure of sensor chip; (d) complete sensor chip
    Mechanical simulation of package structure. (a) Pressure from multiple directions; (b) pressure from single direction
    Vertical displacement of bottom layer of PDMS
    Testing system for performance of R-OMCSC
    Characteristics of R-OMCSC. (a) Pressure sensitivity; (b) voltage change of pressure applied at different locations; (c) response time; (d) repeatability; (e) different detection locations in radial artery; (f) pulse waves at different locations of human pulse
    Typical pulse waveform of brachial artery and radial artery
    Dual-channel pulse wave detection method
    Pulse wave preprocessing process
    Grid search results for c and g. (a) SBP model; (b) DBP model
    Output of SVR prediction model. (a) SBP model; (b) DBP model
    Correlation plots and Bland-Altman plots of dual-channel BP detection model based on SVR. (a)(c) SBP model; (b)(d) DBP model
    Changes of BPand K value. (a) Different times in one day; (b) exercise time
    • Table 1. Measurement range of RPTT, BPTT, and DBRPTT

      View table

      Table 1. Measurement range of RPTT, BPTT, and DBRPTT

      ParameterRPTT /msBPTT /msDBRPTT /msNormalization
      Maximum282270901
      Minimum230217250
    • Table 2. Performance comparison with other blood pressure sensors

      View table

      Table 2. Performance comparison with other blood pressure sensors

      Sensor typeSensor locationPulse pressure sensitivityResponse time /msSBP accuracy (MD, SD)DBP accuracy (MD, SD)Reference
      PPGFinger tips--0.06, 7.080.01, 4.66[29] (2016)
      OIFace--0.39, 7.30-0.20, 6.00[30] (2019)
      OFCDRadial artery--0.24, 2.390.12, 2.62[15] (2020)
      AFMSCRadial artery-0.2211--[16] (2022)
      AFOFDRadial artery-213 μW/kPa5-0.35, 4.68-2.54, 4.07[31] (2023)
      Double R-OMCRadial artery, brachial artery-0.682 kPa-1350.08, 1.13-0.35, 1.25This work
    Tools

    Get Citation

    Copy Citation Text

    Xue Zou, Junhao Fan, Binbin Luo, Fumin Zhou, Decao Wu, Zufan Zhang, Mingfu Zhao. Dual-Channel Reflective Optical Microfiber Coupler Diaphragms for Continuous and Accurate Blood Pressure Monitoring[J]. Acta Optica Sinica, 2024, 44(7): 0728001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Remote Sensing and Sensors

    Received: Nov. 3, 2023

    Accepted: Dec. 13, 2023

    Published Online: Apr. 11, 2024

    The Author Email: Luo Binbin (luobinbin@cqut.edu.cn)

    DOI:10.3788/AOS231738

    CSTR:32393.14.AOS231738

    Topics