Acta Optica Sinica, Volume. 43, Issue 19, 1912001(2023)

Influence of Installation and Adjustment Error of Gravitational Wave Telescope on TTL Noise

Xinxu Cui1, Chao Fang1、*, and Zhi Wang1,2、**
Author Affiliations
  • 1Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033,Jilin , China
  • 2School of Fundamental Physics and Mathematical Sciences, Hangzhou Institute for Advanced Study, University of Chinese Academy of Sciences , Hangzhou 310024, Zhejiang , China
  • show less
    Figures & Tables(10)
    Optical path diagram of space gravitational wave telescope
    Wavefront difference of space gravitational wave telescope
    Optical path diagram of interstellar laser interferometer
    There is an installation error between aperture and primary mirror. (a) ΔSLPF change curve; (b) Line graph of ΔSLPF rate of change
    There is an installation error between primary mirror and secondary mirror. (a) ΔSLPF change curve; (b) Line graph of ΔSLPF rate of change
    There is an installation error between secondary mirror and third mirror. (a) ΔSLPF change curve;(b) Line graph of ΔSLPF rate of change
    There is an installation error between third mirror and fourth mirror. (a) ΔSLPF change curve; (b) Line graph of ΔSLPF rate of change
    Equivalent optical path diagram of gravitational wave telescope
    • Table 1. Structural parameters of off-axis quadruple mirror telescope

      View table

      Table 1. Structural parameters of off-axis quadruple mirror telescope

      No.NameRadius of curvature /mmThickness /mm
      1Primary mirror-12xx.xxx-600
      2Secondary mirror-8x.xxx660
      3Third mirror-6xx.xxx-58.141
      4Forth mirror3xx.xxx118.013
    • Table 2. Distance tolerance values ​​of installation and adjustment

      View table

      Table 2. Distance tolerance values ​​of installation and adjustment

      No.Component distanceDesign distance /mmTolerance value /mmActual distance /mm
      1Aperture and primary mirror600.0000.200600.200
      2Primary and secondary mirror-600.000-0.200-600.200
      3Secondary and third mirror660.0000.200660.200
      4Third and fourth mirror-58.141-0.200-58.341
    Tools

    Get Citation

    Copy Citation Text

    Xinxu Cui, Chao Fang, Zhi Wang. Influence of Installation and Adjustment Error of Gravitational Wave Telescope on TTL Noise[J]. Acta Optica Sinica, 2023, 43(19): 1912001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Mar. 14, 2023

    Accepted: Apr. 17, 2023

    Published Online: Oct. 23, 2023

    The Author Email: Chao Fang (ciompfangchao@126.com), Zhi Wang (wz070611@126.com)

    DOI:10.3788/AOS230675

    Topics