Chinese Optics, Volume. 15, Issue 2, 276(2022)

Non-contact perception of physiological parameters from videos of faces

Xiao-qiang JI*, Zhen-yao LIU, Bing-lin LI, Zhi RAO, Gui-wen LI, and Li-wei SU
Author Affiliations
  • School of Life Science and Technology, Changchun University of Science and Technology, Changchun 130022, China
  • show less
    Figures & Tables(18)
    Flow chart of video image processing
    Comparison of IPPG source signals before and after amplification
    Flow chart of IPPG source signal processing
    The signal of each component after PCA dimensionality reduction
    Independent source signal separated by a blind source
    Pulse wave signal after bandpass filtering
    Fourier transform spectrogram
    Respiration signal after filtering
    Schematic diagram and real diagram of the experimental acquisition device
    Comparison of heart rate estimation results in three scenarios
    Bland-Altman consistency analysis of heart rate results in three scenarios
    Comparison of respiratory rate estimation results in three scenarios
    Bland-Altman consistency analysis of respiratory rate results in three scenarios
    Comparison of SpO2 estimation results in three scenarios
    Bland-Altman consistency analysis of SpO2 results in three scenarios
    • Table 1. Performance comparison of various heart rate detection algorithms under natural light

      View table
      View in Article

      Table 1. Performance comparison of various heart rate detection algorithms under natural light

      方法|Me| (time·min−1) SDe(time·min−1) RMSE(time·min−1) MerCor
      文献[6] 1.451.942.061.95%0.9278
      文献[12] 2.393.563.383.05%
      本文方法1.781.771.982.54%0.9668
    • Table 2. Performance comparison of various respiratory rate detection algorithms under natural light

      View table
      View in Article

      Table 2. Performance comparison of various respiratory rate detection algorithms under natural light

      方法Me(time·min−1) |Me| (time·min−1) SDe(time·min−1) RMSE(time/min−1) Cor
      文献[10] −0.582.543.984.020.61
      本文方法0.631.781.881.980.60
    • Table 3. Performance comparison of SpO2 detection algorithms under natural light

      View table
      View in Article

      Table 3. Performance comparison of SpO2 detection algorithms under natural light

      方法MeSDeRMSEMer
      文献[8] 0.043%1.10%2.02%2.7%
      文献[11] 1.00%1.32%0.87%
      本文方法−0.27%1.05%1.08%0.82%
    Tools

    Get Citation

    Copy Citation Text

    Xiao-qiang JI, Zhen-yao LIU, Bing-lin LI, Zhi RAO, Gui-wen LI, Li-wei SU. Non-contact perception of physiological parameters from videos of faces[J]. Chinese Optics, 2022, 15(2): 276

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Original Article

    Received: Aug. 12, 2021

    Accepted: --

    Published Online: Mar. 28, 2022

    The Author Email:

    DOI:10.37188/CO.2021-0157