Chinese Optics Letters, Volume. 18, Issue 8, 081202(2020)

Hybrid calibration method of a wide-view-angle Mueller polarimeter for hyper-numerical-aperture imaging systems

Jianhui Li, Yanqiu Li*, Ke Liu, Guodong Zhou, Lihui Liu, and Meng Zheng
Author Affiliations
  • Key Laboratory of Photoelectronic Imaging Technology and System of Ministry of Education of China, School of Optics and Photonics, Beijing Institute of Technology, Beijing 100081, China
  • show less
    Figures & Tables(8)
    Schematic diagram of a wide-view-angle Mueller polarimeter for measuring the Mueller pupil of projection optics.
    Experimental setup for measuring Mueller pupil of projection optics: 1, laser; 2, reflector; 3, beam expander; 4, uniformizer; 5, PSG; 6, projection optics; 7, focal plane; 8, collimating optics; 9, PSA; 10, CCD; 11, controller.
    Mueller matrices over two-dimensional distribution of air (a) before calibration and (b) after calibration.
    Mueller matrix over two-dimensional distribution of a QWP with its fast axis at 30° using the calibrated PSG and PSA. (a) The corresponding polarization parameters. (b) d, diattenuation; R, total retardance; Δ, depolarization coefficient; δ, θ, linear retardance and fast-axis orientation; and ψ, circular retardance.
    (a) Product of Mueller pupils of projection optics and collimating optics, M; (b) the Mueller pupil of collimating optics, Mc; (c) the Mueller pupil of projection optics, Mp.
    Comparison of polarization parameters extracted from obtained M, Mc, and Mp shown in Fig. 5.
    • Table 1. Parameter Errors of PSG and PSA

      View table
      View in Article

      Table 1. Parameter Errors of PSG and PSA

      ErrorsMean Value (°)
      ε11.2034
      ε20.8425
      ε3−0.3839
      ε4−1.3407
      ε5−1.5470
    • Table 2. Corresponding Mean Values, RMS, and PV Values of Each Sub-Pupil of Mp and M in Fig. 5

      View table
      View in Article

      Table 2. Corresponding Mean Values, RMS, and PV Values of Each Sub-Pupil of Mp and M in Fig. 5

      MpM
      Mean[1.00000.01750.00100.00290.02350.95780.12660.00050.01280.10621.00770.06070.00400.01890.08050.9770][1.00000.01660.00200.00350.00520.94900.00660.00920.00030.00240.93790.00270.00210.00130.01240.9222]
      RMS[00.03000.02220.00780.03420.04640.13270.05150.04040.11050.05970.10980.01500.04700.11430.0373][00.02250.02160.01030.01370.05070.02550.00450.01250.01980.06160.04450.01060.03900.04810.0777]
      PV[00.69631.08620.18440.92581.93792.28260.42791.05842.43253.73501.31790.43040.51491.06902.7689][00.13330.20790.15300.13140.62150.21490.35780.14620.20950.64330.31390.14310.29850.31640.5632]
    Tools

    Get Citation

    Copy Citation Text

    Jianhui Li, Yanqiu Li, Ke Liu, Guodong Zhou, Lihui Liu, Meng Zheng, "Hybrid calibration method of a wide-view-angle Mueller polarimeter for hyper-numerical-aperture imaging systems," Chin. Opt. Lett. 18, 081202 (2020)

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Optical Sensing, Measurements, and Metrology

    Received: Apr. 22, 2020

    Accepted: Jun. 10, 2020

    Posted: Jun. 10, 2020

    Published Online: Jul. 10, 2020

    The Author Email: Yanqiu Li (liyanqiu@bit.edu.cn)

    DOI:10.3788/COL202018.081202

    Topics