Acta Optica Sinica, Volume. 37, Issue 9, 0904001(2017)

Linear Calibration of Wide Dynamic Range Response of Silicon Detectors Based on Double LEDs Light Source Integrating Sphere

Hui Liu1,2、*, Hongyao Chen1, Xiaolong Si1, and Liming Zhang1
Author Affiliations
  • 1 Key Laboratory of Optical Calibration and Characterization, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • 2 University of Science and Technology of China, Hefei, Anhui 230026, China
  • show less
    Figures & Tables(13)
    Schematic of the Si detectors linearity measurement instrument
    Relationship between detector response current and LED injection current
    Nonlinearity correction factor of Si detector
    Linearity error of Si detector
    Relative spectral radiance of two LEDs
    Relative spectral radiance of the LED with different injection currents
    Schematic of the measurement of spectral drift
    Timing diagram of LEDs switch and data collection
    Uncorrected and corrected nonlinearity factors
    Relative uncertainty of nonlinearity factor
    Relative uncertainty of nonlinearity correct factor
    • Table 1. Measurement data of spectral drift

      View table

      Table 1. Measurement data of spectral drift

      Injection current0.76~5.3 μA5.3~92 μA92 μA~2.5 mA2.5 mA~0.57 A
      Nonlinearity factor (OD: 3)-1.0000361.0001511.00011
      Nonlinearity factor (OD: 2)1.002210.999781.0001001.00018
      Nonlinearity factor (OD: 1)0.998250.999951.0000481.00018
      Nonlinearity factor (OD: 0.1)1.004800.999781.0000151.00016
      Spectral drift ur,W2.69×10-31.11×10-45.17×10-52.86×10-5
    • Table 2. Combined uncertainty of measurement system

      View table

      Table 2. Combined uncertainty of measurement system

      LED injection currentResponse currentSpectral drifturW /10-5Correction factorurCn /10-5Combined uncertaintyur /10-5
      0.76-5.3 μA39.2 pA-0.326 nA2697.05269
      5.3-92 μA0.326-81.6 nA11.13.2311.6
      92 μA-2.5 mA81.6 nA-20.4 μA5.171.435.36
      2.5 mA-0.57 A20.4 μA-10.2 mA2.861.433.20
    Tools

    Get Citation

    Copy Citation Text

    Hui Liu, Hongyao Chen, Xiaolong Si, Liming Zhang. Linear Calibration of Wide Dynamic Range Response of Silicon Detectors Based on Double LEDs Light Source Integrating Sphere[J]. Acta Optica Sinica, 2017, 37(9): 0904001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Detectors

    Received: Apr. 17, 2017

    Accepted: --

    Published Online: Sep. 7, 2018

    The Author Email:

    DOI:10.3788/AOS201737.0904001

    Topics