Chinese Optics Letters, Volume. 16, Issue 7, 071701(2018)
Simultaneous blood flow and oxygenation measurements using an off-the-shelf spectrometer
Fig. 1. Schematic of the proposed system. The system is configured with two lasers, a
Fig. 2. Graphical explanation of the procedure to get the blood flow index (BFI) for each wavelength. One square means one pixel. Time varying intensity values for pixels of 785 and 852 nm were used to get the BFI for each wavelength in unit time.
Fig. 3. Variation of normalized BFI of diffuse correlation spectroscopy, DCS (blue square), 785 nm signal (red dot), and 852 nm signal (green down triangle) of the spectrometer measured on the flow phantom while varying flow rate from 0 to 0.02 mL/s (with 0.01 mL/s of increment) and from 0.02 to 0.06 mL/s (with 0.02 mL/s of increment).
Fig. 4. Variation of oxygenated (red square), deoxygenated (blue circle), and total hemoglobin (green up triangle) concentration to test the feasibility of oxygenation measurement of the system by the blood phantom test. 21%
Fig. 5. Changes of oxygenated (red square for the suggested system and magenta up triangle for broadband NIRS), deoxygenated hemoglobin (blue circle for the suggested system and dark olive diamond for broadband NIRS) concentration (top) and percent variation of blood flow (bottom) for 785 nm (black up triangle with dotted line), 852 nm (red circle with short dashed line), and laser Doppler (blue solid square with solid line) during an arterial arm cuff occlusion (1 min of baseline, 3 min of occlusion with 220 mmHg, and another 3 min after cuff release).