Acta Optica Sinica, Volume. 40, Issue 20, 2001003(2020)

Design and Testing of a Portable Multichannel Reference Light Source

Mengfan Li1,2, Yadong Hu1,2、*, Shanshan Cui1,2, Xinyu Yu1,2, Zhenwei Qiu1,2, Zhuoran Li1,2, Xiangjing Wang1,2, Aiwen Zhang1,2, and Xiaobing Sun1,2
Author Affiliations
  • 1Anhui Institute of Optics and Fine Mechanics, Hefei Institutes of Physical Sciences, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • 2Key Laboratory of Optical Calibration and Characterization, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • show less
    Figures & Tables(11)
    Overall structure and working principle of PMRLS
    Head structure of PMRLS and the effect after installation
    Structure of luminous module
    Relative spectrum data of PMRLS and SMAC at 870 nm waveband
    Schematic of VI conversion module of drive circuit
    Non-stability data of RMRLS
    • Table 1. Technical specification of PMRLSunit:%

      View table

      Table 1. Technical specification of PMRLSunit:%

      ParameterTolerance
      Error of energy adjustment±10
      Non-stability1
      Non-repeatability2
    • Table 2. Parameters of luminous elementsunit:nm

      View table

      Table 2. Parameters of luminous elementsunit:nm

      TypeCenter wavelength of luminous elementBandwidth of luminous elementCenter wavelength of SMACBandwidth of SMAC
      JX5050/4904882649020
      JX5050/56557812055020
      JX5050/6706612067020
      JX5050/8508393787040
      JX5050/9109017291020
    • Table 3. Parameters of SMAC

      View table

      Table 3. Parameters of SMAC

      Center wavelength /nmNon-linearity /%Non-stability /%Calibration uncertainty /%
      490(P)0.450.052.35
      5500.640.022.44
      670(P)0.430.012.28
      870(P)0.370.012.03
      9100.370.011.99
      13800.720.152.34
      1610(P)0.270.061.79
      2250(P)0.080.031.98
    • Table 4. Errors of energy adjustment for PMRLS

      View table

      Table 4. Errors of energy adjustment for PMRLS

      Center wavelength /nm490550670870910138016102250
      Ti,j11731.59714.07752.06852.07012.913846.66955.06145.1
      Si,j12538.79959.47218.27047.57095.214367.76587.15741.1
      Error /%6.882.53-6.892.851.173.76-5.29-6.57
    • Table 5. Non-stability and non-repeatability of PMRLSunit:%

      View table

      Table 5. Non-stability and non-repeatability of PMRLSunit:%

      Center wavelength /nm490550670870910138016102250
      Non-stability0.140.050.080.080.070.070.170.09
      Non-repeatability0.901.291.011.121.180.620.540.69
      Combined uncertainty0.911.291.011.121.180.620.570.70
    Tools

    Get Citation

    Copy Citation Text

    Mengfan Li, Yadong Hu, Shanshan Cui, Xinyu Yu, Zhenwei Qiu, Zhuoran Li, Xiangjing Wang, Aiwen Zhang, Xiaobing Sun. Design and Testing of a Portable Multichannel Reference Light Source[J]. Acta Optica Sinica, 2020, 40(20): 2001003

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Atmospheric Optics and Oceanic Optics

    Received: Apr. 30, 2020

    Accepted: Jul. 3, 2020

    Published Online: Sep. 30, 2020

    The Author Email: Yadong Hu (huyadong@aiofm.ac.cn)

    DOI:10.3788/AOS202040.2001003

    Topics