Acta Photonica Sinica, Volume. 54, Issue 4, 0412002(2025)

Spectral Confocal Displacement Measurement Method Based on SNV Correction of Surface Roughness

Chunyan LI, Wenwen FU*, Jihong LIU, Danlin LI, Shaojie WU, Ninglin WANG, and Kaili REN
Author Affiliations
  • School of Electronic Engineering,Xi'an University of Posts & Telecommunications,Xi'an 710121,China
  • show less
    Figures & Tables(11)
    Schematic diagram of spectral confocal displacement measurement system
    Influence of sample surface roughness on spectral confocal shift measurement
    Variation of reflectivity of received beam caused by roughness
    Displacement measurement error caused by reflectivity change caused by roughness
    Spectral response curve
    Spectral response curve after SVM classification
    Calibration model flow chart
    Surface of carbon steel sample with roughness parameter of 35 nm diagram
    Experimental setup
    • Table 1. Experimental measurement results

      View table
      View in Article

      Table 1. Experimental measurement results

      Roughness parameter

      Rq/nm

      Theoretical

      value z0/mm

      Measured value z/mm

      Corrected measured

      value z'/mm

      Error before correction

      Δz/μm

      Error after

      correction Δz'/μm

      1.017 61.012 71.017 44.90.2
      1.109 71.104 61.109 65.10.1
      1.089 61.084 41.089 35.20.3
      1.095 71.089 41.095 66.30.1
      151.106 71.100 31.106 46.40.3
      0.995 40.989 00.994 56.40.9
      1.060 11.052 61.059 17.51.0
      1.132 71.125 21.131 37.51.4
      1.145 71.137 61.144 48.11.3
      0.998 70.990 60.997 38.11.4
      1.057 51.049 21.055 68.31.9
      1.016 21.007 91.015 18.31.1
      251.133 91.125 21.131 98.72.0
      1.097 71.088 51.095 99.21.8
      1.108 71.099 41.106 49.32.3
      1.050 51.040 91.049 59.61.0
      0.871 20.861 10.869 910.11.3
      1.061 71.051 11.059 610.62.1
      1.138 11.127 21.136 910.91.2
      1.143 41.132 31.142 111.11.3
      351.142 31.130 21.140 912.11.4
      1.016 51.004 11.014 712.41.8
      0.729 60.716 80.726 912.82.7
      0.881 10.864 40.877 216.73.9
    • Table 2. Comparing experimental measurement results

      View table
      View in Article

      Table 2. Comparing experimental measurement results

      Roughness parameter

      Rq/nm

      Direct

      measurement maximum error/μm

      Direct measurement

      average error/μm

      Maximum

      error after correction/μm

      Corrected average error/μm
      104.34.10.40.2
      207.97.61.41.2
      3010.09.81.61.3
    Tools

    Get Citation

    Copy Citation Text

    Chunyan LI, Wenwen FU, Jihong LIU, Danlin LI, Shaojie WU, Ninglin WANG, Kaili REN. Spectral Confocal Displacement Measurement Method Based on SNV Correction of Surface Roughness[J]. Acta Photonica Sinica, 2025, 54(4): 0412002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Oct. 13, 2024

    Accepted: Dec. 12, 2024

    Published Online: May. 15, 2025

    The Author Email: Wenwen FU (f17303789391@163.com)

    DOI:10.3788/gzxb20255404.0412002

    Topics