Spectroscopy and Spectral Analysis, Volume. 41, Issue 10, 3021(2021)

Analysis of Metal Impurity Elements in Li4Ti5O12 Through Microwave Plasma Atomic Emission Spectroscopy

Hong-wei LIU1、* and Liang FU2、2; *;
Author Affiliations
  • 11. College of Material and Chemical Engineering, Hunan Institute of Technology, Hengyang 421002, China
  • 22. College of Materials Science and Engineering, Chongqing University, Chongqing 400045, China
  • show less
    Figures & Tables(6)
    Spectral interferences for Pb 405.781 nm corrected using FLIC
    Correction effect of ionization suppression solution CsNO3 on ionization interferences
    Four-hour stability plot for analytes in Li4Ti5O12 sample solution
    • Table 1. Parameter of calibration curve and the method detect limit (MDL)

      View table
      View in Article

      Table 1. Parameter of calibration curve and the method detect limit (MDL)

      ElementLinear range
      /(μg·L-1)
      Coefficient
      (r)
      MDL
      /(μg·g-1)
      ElementLinear range
      /(μg·L-1)
      Coefficient
      (r)
      MDL
      /(μg·g-1)
      Mn0.30~2000.999 70.09Fe2.58~500.999 80.77
      Na0.16~2 0000.999 70.05Al0.70~2000.999 90.21
      Pb1.38~500.999 90.41Mg0.17~2000.999 50.05
      Ni1.03~500.999 80.31Cu1.30~2000.999 30.39
      Cr0.52~500.999 90.16Ca0.27~2000.999 70.08
      Zn0.93~500.999 50.28Cd0.40~500.999 80.12
      K0.10~10 0000.999 80.03Co0.82~500.999 60.25
    • Table 2. Evaluation of the accuracy and precision of analytical methods (n=6)

      View table
      View in Article

      Table 2. Evaluation of the accuracy and precision of analytical methods (n=6)

      ElementSpiked
      /(μg·L-1)
      Total found
      /(μg·L-1)
      RSD
      /%
      Recovery
      /%
      GB/T 30836—2014
      /(μg·L-1)
      t test
      Mn10098.7±2.602.6398.799.1±3.25p=0.41
      Na1 000966±21.22.1996.6974±36.8p=0.33
      Pb109.73±0.282.8897.39.96±0.18p=0.06
      Ni109.91±0.272.7299.110.2±0.33p=0.06
      Cr1010.2±0.343.3310210.3±0.30p=0.30
      Zn109.80±0.191.9498.010.1±0.32p=0.04
      K1 0001030±40.13.891031 060±53.6p=0.15
      Fe1010.2±0.323.1310210.4±0.26p=0.13
      Al100101±3.063.03101102±2.52p=0.28
      Mg100104±2.712.61104102±3.75p=0.16
      Cu10098.2±2.552.6098.2101±2.98p=0.06
      Ca10099.0±3.143.1799.096.6±3.69p=0.13
      Cd1010.3±0.292.8210310.1±0.24p=0.11
      Co109.64±0.171.7696.49.81±0.35p=0.15
    • Table 3. Analytical results of Li4Ti5O12 samples (μg·g-1, n=6)

      View table
      View in Article

      Table 3. Analytical results of Li4Ti5O12 samples (μg·g-1, n=6)

      ElementSample ASample BElementSample ASample B
      FoundCertifiedFoundCertifiedFoundCertifiedFoundCertified
      Mn16.7±0.5116.439.1±1.1038.7Fe7.32±0.307.5012.8±0.4313.1
      Na659±21.0665287±10.4293Al14.3±0.4614.127.1±0.7328.0
      Pb9.22±0.359.434.11±0.184.17Mg38.2±1.0739.521.5±0.6222.1
      Ni1.13±0.021.101.08±0.032.11Cu11.6±0.4111.928.3±0.7929.4
      Cr6.27±0.196.324.31±0.144.28Ca20.8±0.7321.216.0±0.5115.7
      Zn5.13±0.165.082.08±0.072.11Cd1.34±0.031.332.09±0.052.12
      K4 160±16541902 530±90.22 600Co4.67±0.154.722.36±0.042.35
    Tools

    Get Citation

    Copy Citation Text

    Hong-wei LIU, Liang FU. Analysis of Metal Impurity Elements in Li4Ti5O12 Through Microwave Plasma Atomic Emission Spectroscopy[J]. Spectroscopy and Spectral Analysis, 2021, 41(10): 3021

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Sep. 7, 2020

    Accepted: --

    Published Online: Oct. 29, 2021

    The Author Email: Hong-wei LIU (hwl0466@163.com)

    DOI:10.3964/j.issn.1000-0593(2021)10-3021-05

    Topics