Laser & Optoelectronics Progress, Volume. 55, Issue 1, 11201(2018)

Two-Step Calibration of Probe Tip Center of Planar Target

Zheng Xingchun, Zhao Min*, and Feng Shaohua
Author Affiliations
  • School of Mechanical and Precision Instrument Engineering, Xi''an University of Technology,Xi''an, Shaanxi 710048, China
  • show less
    Figures & Tables(9)
    Vision coordinate measurement system of planar target
    Diagram of target in different positions
    Simulation results of coordinate T'x and T'y of probe tip center before and after calibration. (a) T'x; (b) T'y
    Simulation results of corrected value Δx, Δy and Δz of probe tip center
    Simulation results of corrected values Δx and Δy of probe tip center
    Relationship between coordinate T″y of probe tip center and angle tilt β of target
    Trend of coordinate T″y of probe tip center
    • Table 1. Coordinates of targets under target coordinate systemmm

      View table

      Table 1. Coordinates of targets under target coordinate systemmm

      Coordinate location12345678
      x85135175185235285325335
      y00-500050
      z00000000
    • Table 2. Results of probe tip center calibration with the first set of imagesmm

      View table

      Table 2. Results of probe tip center calibration with the first set of imagesmm

      Image numberkProposed methodRef.[8]
      T'xT'yT'zT'xT'yT'z
      1-16.399280.0071931.306-16.495279.8771931.014
      2-16.298280.0401931.864-16.384279.9241931.567
      3-16.300279.9981931.627-16.402279.8891931.327
      4-16.246279.9951931.548-16.368279.8951931.247
      5-16.357279.9921931.567-16.480279.9041931.265
      6-16.429280.0241932.023-16.570279.9501931.717
      7-16.419280.0471931.301-16.577279.9951931.000
      8-16.330280.0781931.769-16.514280.0361931.463
      9-16.303280.0871931.853-16.491280.0571931.545
      10-16.211280.0531932.106-16.407280.0181931.797
      11-16.213280.0771932.156-16.409280.0441931.846
      Mean-16.319280.0361931.738-16.463279.9631931.435
      Standard deviation0.0770.0350.2960.0740.0690.292
    Tools

    Get Citation

    Copy Citation Text

    Zheng Xingchun, Zhao Min, Feng Shaohua. Two-Step Calibration of Probe Tip Center of Planar Target[J]. Laser & Optoelectronics Progress, 2018, 55(1): 11201

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Aug. 1, 2017

    Accepted: --

    Published Online: Sep. 10, 2018

    The Author Email: Zhao Min (zhaomin1973@xaut.edu.cn)

    DOI:10.3788/LOP55.011201