Acta Optica Sinica, Volume. 43, Issue 11, 1117001(2023)

Comparative Analysis of the Performance of PMT and APD Detectors in Confocal Laser Endomicroscopy

Guanting Hou1, Baoteng Xu1,2, Wei Zhou2, Jialin Liu2, Liquan Guo2, and Xibin Yang1,2、*
Author Affiliations
  • 1Division of Life Sciences and Medicine, School of Biomedical Engineering (Suzhou), University of Science and Technology of China, Suzhou 215163, Jiangsu, China
  • 2Center of Light and Health, Suzhou Institute of Biomedical Engineering and Technology, Chinese Academy of Sciences, Suzhou 215163, Jiangsu, China
  • show less
    Figures & Tables(14)
    Schematic of dual optical path CLE system
    Acquisition process of confocal images
    Schematic of image signal-to-noise calculation
    Detector performance comparison experiment device
    RSN_PMT of PMTSS varying with optical power
    Comparison of detector performance. (a) RSN_APD of APD120A2 varying with optical power; (b) RSN of different detectors varying with optical power
    Experimental results of uniform fluorescent solution imaging of APD120A2 and PMTSS
    Imaging results of fluorescent samples and statistics results of the number of pixel points under different optical powers.(a)-(c) Fluorescent beads; (d)-(f) lens tissue; (g)-(i) upper epidermis of Photinia serrulata leaves; (j)-(l) lower epidermis of Photinia serrulata leaves
    Summary of RSN_Pic for imaging fluorescent samples. (a) Boxplot of image SNR; (b) relationship between image SNR and imaging average optical power
    • Table 1. Configuration and parameters of detectors

      View table

      Table 1. Configuration and parameters of detectors

      ParameterPMTAPD
      DetectorPMTSSAPD120A2
      Price(CNY)28663.9111944.83
      Active area diameter3.7 mm×13.0 mmΦ1 mm
      Bandwidth /MHz25050
      Dark current /nA2(Typical);10(Max)
      NEP /(pW·Hz-1/20.21(DC-50 MHz)
      AmplifierC9999(50 mV/μA)DHPVA-101(20 dB)
    • Table 2. Data collection conditions of detector performance comparison experiment

      View table

      Table 2. Data collection conditions of detector performance comparison experiment

      ConditionPMTAPD
      DetectorPMTSSAPD120A2
      Optical powerStart at 5 nW,end at 50 nW,step size is 5 nW
      Sample rate125 MHz
      Gain voltage0.313 V,0.342 V,0.374 Vfixed
    • Table 3. Parameter values used for the SNR calculation

      View table

      Table 3. Parameter values used for the SNR calculation

      ParameterValueParameterValue
      η24%(PMTSS)VAmpPMT2.6 mV(RMS)
      e1.6×10-19 CVAmpAPD283 μV(RMS)
      λ525 nmPNE0.21 pW/Hz
      h

      6.62607015×

      10-34 J·s

      Giv50000
      c3×108 m/sGmodule50000
      B10 MHzGv10
      Id10 nARs22.3 A/W
      Ma50FMa0.4
    • Table 4. Uniform fluorescent solution experimental conditions

      View table

      Table 4. Uniform fluorescent solution experimental conditions

      ConditionPMTAPD
      DetectorPMTSSAPD120A2
      Solution75 mL fluorescein sodium solution with mass concentration of 0.0067%,0.0133%,0.02%,0.0267%,and 0.0333% g/mL
      Gain voltage0.342 Vfixed
      Sample rate125 MHz
      Amount of dataAcquire 120 consecutive images for each condition
    • Table 5. Fluorescence sample selection and preparation

      View table

      Table 5. Fluorescence sample selection and preparation

      No.SamplePreparation
      S1Lower epidermis of Photinia serrulata leavesAvoid the main veins and cut out a number of small square leaves,soak them in 1% sodium fluorescein solution for 96 h,rinse and dry the epidermis before the experiment
      S2Upper epidermis of Photinia serrulata leavesAvoid the main veins and cut out a number of small square leaves,soak them in 1% sodium fluorescein solution for 72 h,rinse and dry the epidermis before the experiment
      S3Lens tissue0.0333% sodium fluorescein solution sprayed and dried
      S413 μm fluorescent beadsPipette 5 μL of fluorescent beads(FL-PS-G-13,TJDAE technology)onto the glass slide and dry
    Tools

    Get Citation

    Copy Citation Text

    Guanting Hou, Baoteng Xu, Wei Zhou, Jialin Liu, Liquan Guo, Xibin Yang. Comparative Analysis of the Performance of PMT and APD Detectors in Confocal Laser Endomicroscopy[J]. Acta Optica Sinica, 2023, 43(11): 1117001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Medical optics and biotechnology

    Received: Dec. 13, 2022

    Accepted: Feb. 9, 2023

    Published Online: Jun. 13, 2023

    The Author Email: Xibin Yang (yangxb@sibet.ac.cn)

    DOI:10.3788/AOS222137

    Topics