Chinese Journal of Lasers, Volume. 50, Issue 21, 2107203(2023)

Quantitative Analysis of Hemoglobin Based on SiPLS‑BP Model

Zhushanying Zhang1,2,3, Sicong Zhu1,2,3, Xianwen Zhang4, Baorong Fu5、*, Zhi Li1,2,3, Huimin Cao1,2,3、**, and Yi Liu3,6
Author Affiliations
  • 1College of Biomedical Engineering, South-Central Minzu University, Wuhan 430074, Hubei, China
  • 2Key Laboratory of Cognitive Science, State Ethnic Affairs Commission, Wuhan 430074, Hubei, China
  • 3Hubei Key Laboratory of Medical Information Analysis and Tumor Diagnosis&Treatment, Wuhan 430074, Hubei, China
  • 4Linyi GREPO Garden Machinery Co., Ltd., Linyi 276700, Shandong, China
  • 5Wuhan Great Sea Hi-Tech Co., Ltd., Wuhan 430223, Hubei, China
  • 6School of Mechanical and Electronic Engineering, Wuhan University of Technology, Wuhan 430070, Hubei, China
  • show less
    Figures & Tables(10)
    Original spectra of imitation solution samples
    Analysis of modeling results of imitation solution samples in different intervals
    Correlation between predicted value and actual value of blood sample under BP model
    Correlation between predicted value and actual value of imitation solution sample under BP model
    Correlation between predicted value and actual value of blood sample under SiPLS-BP model
    Correlation between predicted value and actual value of imitation solution sample under SiPLS-BP model
    • Table 1. Quantitative analysis results of optimal bands and their combinations of imitation solution samples

      View table

      Table 1. Quantitative analysis results of optimal bands and their combinations of imitation solution samples

      Interval No.RcRMSECRpRMSEP
      x30.79396.68380.90375.1108
      x40.86175.63690.96653.2154
      x50.88455.26620.96683.0444
      x60.86935.62040.85255.6042
      x3,x40.84725.66930.98643.3397
      x3,x50.85255.62260.97923.1272
      x3,x60.84175.9640.98342.2128
      x4,x50.86055.37420.95554.7763
      x4,x60.84395.92760.99121.8211
      x5,x60.8745.53080.99593.1854
      x3,x4,x50.85375.61720.9783.288
      x3,x4,x60.85465.81630.97022.8254
      x3,x5,x60.85575.76460.9852.2899
      x4,x5,x60.91314.50940.98072.7211
      x3,x4,x5,x60.85645.80470.9861.9894
      Full-spectrum PLS model0.88575.3670.9573.2282
    • Table 2. Comparison of results of blood samples under different models

      View table

      Table 2. Comparison of results of blood samples under different models

      ModelRcRMSECRpRMSEP
      PLS0.98972.5170.97922.746
      SiPLS0.99371.9680.98941.947
      SPA-PLS0.98802.7170.98432.376
      SiPLS-SPA-PLS0.99361.9920.99061.846
      Full-spectrum BP0.98273.3920.97423.066
      SiPLS-BP0.98652.9100.99071.807
    • Table 3. Comparison of results of imitation solution samples under different models

      View table

      Table 3. Comparison of results of imitation solution samples under different models

      ModelRcRMSECRpRMSEP
      PLS0.88575.3670.9573.228
      SiPLS0.84395.9280.99121.821
      Full-spectrum BP12.765×10-100.99151.554
      SiPLS-BP17.305×10--90.99751.017
    • Table 4. Screening bands of blood samples under different models

      View table

      Table 4. Screening bands of blood samples under different models

      ModelBand range /nm
      SiPLS1100‒1298,1600‒1798,2100‒2198
      SiPLS-BP1100‒1310,1630‒1840,2054‒2370
    Tools

    Get Citation

    Copy Citation Text

    Zhushanying Zhang, Sicong Zhu, Xianwen Zhang, Baorong Fu, Zhi Li, Huimin Cao, Yi Liu. Quantitative Analysis of Hemoglobin Based on SiPLS‑BP Model[J]. Chinese Journal of Lasers, 2023, 50(21): 2107203

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Optical Diagnostics and Therapy

    Received: May. 18, 2023

    Accepted: Jul. 17, 2023

    Published Online: Nov. 1, 2023

    The Author Email: Baorong Fu (fbr1982@163.com), Huimin Cao (huimincao@mail.scuec.edu.cn)

    DOI:10.3788/CJL230845