Photonics Research, Volume. 9, Issue 6, 899(2021)
Single-sweep volumetric optoacoustic tomography of whole mice
Fig. 1. Single-sweep volumetric optoacoustic tomography (sSVOT) system characterization. (a) Schematic of the sSVOT scanner showing the difference between the single-beam illumination based (left) and multibeam illumination (right) approach. SA, spherical array; FB, fiber bundle; and OA, optoacoustic. (b) Simulated light distribution models for single-beam illumination (left) and multibeam illumination (right). (c) Maximum intensity projections (MIPs) across cross-sectional view demonstrating the spheres using single-beam (top left) and multibeam illumination (top right) approaches at single position of the spherical array. The corresponding fluence corrected images are shown at the bottom row. Arrows point to the spheres that appeared after the fluence correction. (d) Characterization of the reconstructed microsphere size in the central imaging plane along the radial (
Fig. 2.
Fig. 3. sSVOT images acquired from different viewing angles (from left to right: front, left back, back, right back) at a 10 mm/s scan speed (6.9 s total scan time per compounded image) with the 16× subelements and icmax compounding method: 1, brown adipose tissue; 2, spinal cord; 3, spleen; 4, kidney; 5, liver; 6, cecum; 7, heart; 8, duodenum; and 9, thoracic vessels. Scale bar: 1 cm.
Fig. 4. Cross-sectional image quality improvement with multibeam illumination for full rotation acquisitions. (a) Schematic set up (top view) for the full (360°) rotation of the spherical array using single-beam (left) and multibeam illumination (right). (b) Corresponding cross-sectional MIP images reconstructed over a 3 mm thickness at various elevational anatomical positions. Scale bar: 1 cm.
Fig. 5. Different reconstruction methods using subelement based back-projection algorithm. (a) Illustration of the spherical array with 512 sensing elements. (b) Subelement divisions used by the reconstruction algorithm are shown in zoom-ins:
Fig. 6. Performance comparison of sSVOT reconstruction performed with different compounding methods: summation (sum), inverse center distance weighting (ICDW), maximum (max), sum with weighted max (sumax), and ICDW with weighted max (icmax). (a) sSVOT reconstructed image using icmax compounding method. Scale bar: 1 cm. (b) Zoomed-in regions of interest (ROI1, ROI2, and ROI3) compare the differences when employing various volume compounding techniques. Arrows point to the differences.
Fig. 7. Performance comparison of sSVOT system for different scan velocities of 10, 20, 40, and 80 mm/s and subelement-based reconstructions. (a) Reconstructed mice volume for a single vertical sweep at a 80 mm/s scan speed using
Get Citation
Copy Citation Text
Sandeep Kumar Kalva, Xose Luis Dean-Ben, Daniel Razansky, "Single-sweep volumetric optoacoustic tomography of whole mice," Photonics Res. 9, 899 (2021)
Category: Imaging Systems, Microscopy, and Displays
Received: Jan. 4, 2021
Accepted: Mar. 15, 2021
Published Online: May. 7, 2021
The Author Email: Daniel Razansky (daniel.razansky@uzh.ch)