Chinese Journal of Lasers, Volume. 47, Issue 8, 804003(2020)

Detection of Solution Refractive Index Variation Based on Optical Fiber Surface Plasmon Resonance

Chen Qianghua1、*, Han Wenyuan1, Kong Xiangyue1, He Yongxi2, and Luo Huifu2
Author Affiliations
  • 1School of Mechanical and Materials Engineering, North China University of Technology, Beijing 100041, China
  • 2School of Mechanical Engineering, Beijing Institute of Technology, Beijing 100081, China
  • show less
    Figures & Tables(9)
    Structure diagram of optical fiber SPR sensor
    Relationship between reflectivity and incident angle. (a) Relationship between reflectivity of p and s components and incident angle; (b) relationship between reflectivity of p component and incident angle at different thicknesses of gold film
    Relationship of p and s components' relevant parameters. (a) Relationship between phase and incident angle; (b) relationship between phase difference and refractive index
    Non-linear system error of optical fiber SPR sensor
    Schematic of optical fiber SPR sensing system
    Experimental curve of measurement stability of proposed system
    Calibration of phase response of glycerine solution
    • Table 1. Measured phase difference data

      View table

      Table 1. Measured phase difference data

      Mass fraction /%Average phasedifference /(°)Standarddeviation /(°)
      087.6120.134
      0.187.6110.162
      0.586.4720.147
      185.3790.138
      1.584.2740.129
      283.1570.132
    • Table 2. Measurement comparison of refractive index of glycerine solution

      View table

      Table 2. Measurement comparison of refractive index of glycerine solution

      Massfraction /%Average phasedifference /(°)n-1.333Refractive index error betweentwo methods /10-5
      Proposed methodMethod in Ref.[22]
      185.3790.0013660.00143.4
      1.584.2740.0020410.00215.9
      283.1570.0027240.00287.6
    Tools

    Get Citation

    Copy Citation Text

    Chen Qianghua, Han Wenyuan, Kong Xiangyue, He Yongxi, Luo Huifu. Detection of Solution Refractive Index Variation Based on Optical Fiber Surface Plasmon Resonance[J]. Chinese Journal of Lasers, 2020, 47(8): 804003

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Measurement and metrology

    Received: Feb. 19, 2020

    Accepted: --

    Published Online: Aug. 17, 2020

    The Author Email: Qianghua Chen (chenqianghua@tsinghua.org.cn)

    DOI:10.3788/CJL202047.0804003