Chinese Journal of Lasers, Volume. 47, Issue 8, 804001(2020)
Method for Measurement Range Extension of Cavity Ring-Down Spectroscopy Based on Threshold Modification
Fig. 1. Influence of setting sampling threshold on detection accuracy under different decay time and relationship between detection accuracy and ring-down time when threshold is fixed. (a) Relationship between detection accuracy and threshold; (b) relationship between detection accuracy and ring-down time when VT=80 mV
Fig. 2. Setting threshold and ring-down time versus methane detection range. (a) Setting threshold versus methane detection range; (b) ring-down time versus methane detection range
Fig. 5. Thresholds and ring-down time under different methane concentrations. (a) Thresholds under different methane concentrations; (b) ring-down curve of methane standard gas with volume fraction of 1.0×10-6 when τ=24.57 μs and VT=238.54 mV; (c) ring-down curve of methane standard gas with volume fraction of 1.0×10-4 when τ=1.36 μs and VT=5.36 mV
Get Citation
Copy Citation Text
Mo Zeqiang, Yu Jin, He Jianguo, Wang Jinduo, Liu Yang, Dai Shoujun, Wang Xiaodong, Meng Jingjing, Xu Yuyang. Method for Measurement Range Extension of Cavity Ring-Down Spectroscopy Based on Threshold Modification[J]. Chinese Journal of Lasers, 2020, 47(8): 804001
Category: Measurement and metrology
Received: Jan. 18, 2020
Accepted: --
Published Online: Aug. 17, 2020
The Author Email: Jin Yu (jinyu@aoe.ac.cn)