Chinese Journal of Lasers, Volume. 47, Issue 2, 207042(2020)

Measurement of Heart Rate Based on Ballistocardiography

Kong Lingqin1, Wu Yuheng1, Pang Zongguang2, Zhao Yuejin1、*, Dong Liquan1, Liu Ming1, Hui Mei1, Wang Weijie1, Guo Ying1, and Wang Xiatian1
Author Affiliations
  • 1Beijing Key Laboratory for Precision Optoelectronic Measurement Instrument and Technology,School of Optics and Photonics, Beijing Institute of Technology, Beijing 100081, China
  • 2North Information Control Research Academy Group Co., Ltd., Nanjing, 211153, China
  • show less
    Figures & Tables(13)
    Involuntarily periodic oscillation in BCG
    OLA-BCG principle. (a) OLA-BCG experimental device diagram; (b) OLA-BCG schematic diagram
    Feature points selection. (a) Feature points selection in Ref.[11]; (b) feature points selection in Ref.[12]; (c) feature points selection in Ref.[17]; (d) reflected laser feature point in OLA-BCG
    Algorithm flowchart
    Calculation flow chart of centroid of reflection spot
    Original BCG signals in different directions. (a) x direction; (b) y direction
    Frequency domain diagrams of original BCG signals in different directions. (a) x direction; (b) y direction
    BCG signal filtered by ICA
    Schematic diagram of SNR calculation
    Bland-Altman analysis result and Pearson correlation coefficient. (a) Bland-Altman analysis result; (b) Pearson correlation coefficient
    • Table 1. Summary of evaluation indicators

      View table

      Table 1. Summary of evaluation indicators

      IndicatorDescriptionCalculation formula
      Mean_differenceMean error of heart rate differenceMean_difference=mean(Hr_difference)
      Sd_differenceStandard deviation of heart rate differenceSd_difference=sqrt(Hr_difference)
      ERMSRoot-mean-square error of heart rate differenceERMS=D(Hr_difference)
      Pmax_differenceMaximum value of heart rate differencePmax_difference=max(Hr_difference)
      rPearson correlation coefficient between Hr_BVP and Hr_BCGr=corr(Hr_BVP,Hr_BCG)
      RSNRatio of power of main pulsatile componentand power of background noise[29]RSN=C1C2y(f)dfB1B2y(f)df-C1C2y(f)df
    • Table 2. Evaluation index statistics of different methods

      View table

      Table 2. Evaluation index statistics of different methods

      Evaluation indexMethod in Ref.[11]Method in Ref.[12]Method in Ref.[17]OLA-BCG
      Mean_difference /(beat·min-1)2.581.50.79-0.67
      Sd_difference /(beat·min-1)2.713.353.102.15
      ERMS /(beat·min-1)3.663.563.122.22
      Mean_SNR /dB-5.45-9.42-7.831.70
      Pmax_difference /(beat·min-1)6554
      r0.710.490.890.88
    • Table 3. Improvement of different evaluation indexes

      View table

      Table 3. Improvement of different evaluation indexes

      MethodΔMean_difference /%ΔSd_difference /%ΔERMS /%ΔMean_SNR /dBΔPmax_difference /%Δr /%
      Method in Ref.[11]74.0320.6639.347.1533.3423.94
      Method in Ref.[12]55.3335.8237.6411.1220.0079.59
      Method in Ref.[17]15.1930.6528.859.5320.00-1.12
    Tools

    Get Citation

    Copy Citation Text

    Kong Lingqin, Wu Yuheng, Pang Zongguang, Zhao Yuejin, Dong Liquan, Liu Ming, Hui Mei, Wang Weijie, Guo Ying, Wang Xiatian. Measurement of Heart Rate Based on Ballistocardiography[J]. Chinese Journal of Lasers, 2020, 47(2): 207042

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: biomedical photonics and laser medicine

    Received: Oct. 8, 2019

    Accepted: --

    Published Online: Feb. 21, 2020

    The Author Email: Yuejin Zhao (yjzhao@bit.edu.cn)

    DOI:10.3788/CJL202047.0207042

    Topics