Acta Optica Sinica, Volume. 42, Issue 4, 0404001(2022)

Responsivity Calibration Technology for Pyroelectric Terahertz Detector

Bin Wu1,2、*, Hongyuan Liu2, Yixin Wang3, Hongchao Wang2, Qiantao Cao2, Peng Zhang2, and Chengping Ying2
Author Affiliations
  • 1Science and Technology on Electronic Test & Measurement Laboratory, Qingdao, Shandong 266555, China;
  • 2The 41st Institute of China Electronics Technology Group Corporation, Qingdao, Shandong 266555, China
  • 3International College of Chongqing University of Posts and Telecommunications, Chongqing 400065, China
  • show less
    Figures & Tables(5)
    Schematic diagram of responsivity calibration system for terahertz detector
    • Table 1. Uncertainty analysis of responsivity calibration system

      View table

      Table 1. Uncertainty analysis of responsivity calibration system

      Source of uncertaintyUncertaintyUncertainty typeProbability distribution
      Responsivity of standard detectoru1=2.0%BUniform
      Output measurement of DUTu2=9.0×10-5BUniform
      Output measurement of standard detectoru3=9.0×10-5BUniform
      Stability of terahertz sourceu4=1.2%BUniform
      Environment noiseu5=0.43%AUniform
    • Table 2. Comparison of terahertz detector calibration research works

      View table

      Table 2. Comparison of terahertz detector calibration research works

      InstitutionCalibrationfrequency /THzUncertainty of powerstandard (k=1)Uncertainty ofcalibration system (k=1)Universality
      PTB1.00--5.001.6%2.0%Good
      NIST0.76, 2.523.3%
      NMIJ0.30, 1.00, 1.502.1%
      NIM0.60--0.90, 2.521.6%2.0%Good
      Our group0.84--7.002.0%2.4%Good
    • Table 3. Calibration data of 12D-3S-VP type terahertz detector

      View table

      Table 3. Calibration data of 12D-3S-VP type terahertz detector

      TimeUNsUsUNdUdRd,corrected
      10.0061.4570.0051.052197.999
      20.0061.4210.0071.024197.219
      30.0041.3850.0050.996196.908
      40.0051.4110.0041.031200.433
      50.0061.4220.0061.028198.049
      60.0051.3860.0050.987195.120
    • Table 4. Calibration data of 6342A type terahertz detector

      View table

      Table 4. Calibration data of 6342A type terahertz detector

      TimeUNsUsUNdUdRd,corrected
      10.0051.4330.0061.876359.333
      20.0051.4130.0041.833356.447
      30.0041.3960.0041.854364.684
      40.0051.3930.0061.844363.363
      50.0071.4170.0051.855360.028
      60.0061.3930.0041.871369.362
    Tools

    Get Citation

    Copy Citation Text

    Bin Wu, Hongyuan Liu, Yixin Wang, Hongchao Wang, Qiantao Cao, Peng Zhang, Chengping Ying. Responsivity Calibration Technology for Pyroelectric Terahertz Detector[J]. Acta Optica Sinica, 2022, 42(4): 0404001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Detectors

    Received: Jul. 7, 2021

    Accepted: Aug. 20, 2021

    Published Online: Jan. 29, 2022

    The Author Email: Wu Bin (wubinw@126.com)

    DOI:10.3788/AOS202242.0404001

    Topics