Chinese Journal of Lasers, Volume. 45, Issue 7, 0711003(2018)

Signal Stability of Collinear Double Pulse Laser Induced Breakdown Spectroscopy Combining with Auto-Focusing System

Jin Wu1,2, Yuejin Wu1, Shuang Fan1,2, Zhuopin Xu1,2, Huiru Chen1,2, Cong Xu1,2, and Qi Wang、* *
Author Affiliations
  • 1 Institute of Technical Biology and Agriculture Engineering, Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • 2 University of Science and Technology of China, Hefei, Anhui 230026, China
  • show less
    Figures & Tables(9)
    Schematic of collinear DP-LIBS experimental system
    Image acquisition versus displacement of z axis
    Process of auto-focusing. (a) The first search process; (b) the second search process
    Relationship between pulse interval and spectral intensity. (a) Atomic lines; (b) ion lines
    Relative standard deviation of spectralintensity versus spectral line
    Calibration curves of Cr element at two modes. (a) Without auto-focusing system; (b) with auto-focusing system
    Calibration curves of Mn element at two modes. (a) Without auto-focusing system; (b) with auto-focusing system
    • Table 1. Sharpness evaluation function value

      View table

      Table 1. Sharpness evaluation function value

      Numberabcde
      Value278577926640327
    • Table 2. Relative standard deviation (RSD) of spectral intensity with and without auto-focusing system

      View table

      Table 2. Relative standard deviation (RSD) of spectral intensity with and without auto-focusing system

      ElementFe INi IFe IICr IFe IFe IICr IINi INi II
      Wavelength /nm358.12352.45238.20357.87373.49238.20206.15361.94218.55
      RSD without auto-focusing0.1380.1280.2040.1870.1390.2030.1320.1240.193
      RSD with auto-focusing0.0520.0580.0840.0690.0500.0840.0640.0550.082
    Tools

    Get Citation

    Copy Citation Text

    Jin Wu, Yuejin Wu, Shuang Fan, Zhuopin Xu, Huiru Chen, Cong Xu, Qi Wang. Signal Stability of Collinear Double Pulse Laser Induced Breakdown Spectroscopy Combining with Auto-Focusing System[J]. Chinese Journal of Lasers, 2018, 45(7): 0711003

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: spectroscopy

    Received: Feb. 5, 2018

    Accepted: --

    Published Online: Sep. 11, 2018

    The Author Email: Qi Wang (wangqi@ipp.ac.cn)

    DOI:10.3788/CJL201845.0711003

    Topics