Acta Optica Sinica, Volume. 44, Issue 18, 1822002(2024)

Optical System Design of UV-Visible Reflective Imaging Micro-Spectrometer

Baoze Guo1,2,3,4, Entao Shi1,3,4、*, Yongmei Wang1,3,4, Jiu Liu1,3,4, Jun Zhu5, and Shun Yao5
Author Affiliations
  • 1National Space Science Center, Chinese Academy of Sciences, Beijing 100190, China
  • 2North China Research Institute of Electro-Optics, Beijing 100015, China
  • 3Beijing Key Laboratory of Space Environment Exploration, Beijing 100190, China
  • 4Key Laboratory of Space and Technology on Space Environmental Situation Awareness, Chinese Academy of Sciences, Beijing 100190, China
  • 5DFH Satellite Co., Ltd., Beijing 100094, China
  • show less
    Figures & Tables(16)
    Diagram of finite coaxial four-mirror optical structure
    Diffraction angle of diffraction grating
    Diagram of spectral system structure
    Diagram of chief ray of on-axis system
    Optical layouts of reflection system. (a) Image plane is away from the optical axis; (b) image plane towards the optical axis
    Diagram of off-axis structure
    Diagram of unshielded structure
    Flow chart of calculation of initial structure
    Optical structure of reflecting spectrometer
    Grayscale images of system wavefront error. (a) 320 nm; (b) 500 nm
    RMS and MTF curves of system varying with wavelength. (a) RMS; (b) MTF
    Smile varying with field of view and Keystone varying with wavelength. (a) Smile varying with field of view; (b) Keystone varying with wavelength
    Spectral resolution. (a) 410 nm; (b) 320-500 nm
    • Table 1. Parameter value ranges

      View table

      Table 1. Parameter value ranges

      ParameterValue
      d1[30, 85]
      d2[30, 60]
      d31[40, 80]
      d32[100, 150]
      d4[130, 200]
      r1≤400
      r2≤200
      r3≤200
      r4≥-100
    • Table 2. Main parameters

      View table

      Table 2. Main parameters

      Performance parameterValue
      Spectral range /nm320-500
      Volume /L3
      Slit length /mm12
      Slit width /mm0.1
      Spectral resolution /nm0.5
      Scaling factor1/2
    • Table 3. Initial structure parameters of spectrometer system

      View table

      Table 3. Initial structure parameters of spectrometer system

      Parameter

      Radius /

      mm

      Distance /

      mm

      Tilt /(°)Line density /μm
      Object-35.1000
      M1 (stop)14.78-33.5600
      M279.3272.1920
      Grating--112.42002.151.458
      M3138.24177.6174
      M4-29.91-17.9000
    Tools

    Get Citation

    Copy Citation Text

    Baoze Guo, Entao Shi, Yongmei Wang, Jiu Liu, Jun Zhu, Shun Yao. Optical System Design of UV-Visible Reflective Imaging Micro-Spectrometer[J]. Acta Optica Sinica, 2024, 44(18): 1822002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Optical Design and Fabrication

    Received: Nov. 20, 2023

    Accepted: Feb. 26, 2024

    Published Online: Sep. 11, 2024

    The Author Email: Shi Entao (set@nssc.ac.cn)

    DOI:10.3788/AOS231807

    CSTR:32393.14.AOS231807

    Topics