NUCLEAR TECHNIQUES, Volume. 46, Issue 6, 060402(2023)

Influence of interference signals on spectrometer and measurement results in positron annihilation lifetime measurements

Lei WANG1,2, Xionggang PENG2,3, Hongqiang ZHANG2,3, Runsheng YU2,3, Peng ZHANG2, Haibiao WU1, Yaru WU2, Baoyi WANG2,3, Xingzhong CAO2,3, Detao XIAO1、*, and Fuyan LIU2、**
Author Affiliations
  • 1School of Nuclear Science and Technology, University of South China, Hengyang 421001, China
  • 2Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049, China
  • 3University of Chinese Academy of Sciences, Beijing 100049, China
  • show less
    Figures & Tables(7)
    Principal scheme of the positron annihilation lifetime spectrometer
    Positron-lifetime spectrum of pure iron(Fe) in 60Co/22Na with activity radio of 3.3 (color online)
    Positron-lifetime spectrum of irradiated RPV steel samples (color online)
    FWHM and FWTM parameter of positron-lifetime spectrum
    • Table 1. Resolved results of positron-lifetime spectrum

      View table
      View in Article

      Table 1. Resolved results of positron-lifetime spectrum

      名称 Typeτ1 / nsI1 / %τ2 / nsI2 / %χ2时间分辨 Time resolution / ps
      标准谱仪 Standard spectrometer0.11098.81.5881.21.08213
      搭建谱仪 Experimental spectrometer0.11097.81.8052.20.97198
    • Table 2. Resolved results of pure iron(Fe) positron-lifetime spectrum

      View table
      View in Article

      Table 2. Resolved results of pure iron(Fe) positron-lifetime spectrum

      名称 Typeτ1 / nsI1 / %τ2 / nsI2 / %τ3 / nsI3 / %χ2PVR/PVRNa
      Fe0.11072.80.38225.61.8051.60.971
      Cs/Na 2.20.10974.10.38224.91.7461.00.920.60
      Co/Na 1.90.10378.30.36920.01.5631.71.080.19
      Co/Na 3.30.08982.50.34316.01.5701.51.030.13
    • Table 3. Resolved results of RPV steel positron-lifetime spectrum

      View table
      View in Article

      Table 3. Resolved results of RPV steel positron-lifetime spectrum

      样品名称 Sampleτ1 / nsI1 / %τ2 / nsI2 / %τ3 / nsI3 / %χ2PVR/PVRNaACo-60 / MBq

      未辐照RPV钢

      Non-irradiated RPV

      0.14597.41.9552.61.0410

      辐照后RPV(0.5 mm)钢

      Irradiated RPV (0.5 mm)

      0.12059.60.26339.00.8801.41.010.311.5

      辐照后RPV(1 mm)钢

      Irradiated RPV (1 mm)

      0.07847.10.21649.70.7043.21.080.182.8
    Tools

    Get Citation

    Copy Citation Text

    Lei WANG, Xionggang PENG, Hongqiang ZHANG, Runsheng YU, Peng ZHANG, Haibiao WU, Yaru WU, Baoyi WANG, Xingzhong CAO, Detao XIAO, Fuyan LIU. Influence of interference signals on spectrometer and measurement results in positron annihilation lifetime measurements[J]. NUCLEAR TECHNIQUES, 2023, 46(6): 060402

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Mar. 6, 2023

    Accepted: --

    Published Online: Jul. 5, 2023

    The Author Email:

    DOI:10.11889/j.0253-3219.2023.hjs.46.060402

    Topics