Acta Optica Sinica, Volume. 44, Issue 14, 1404001(2024)

Error Analysis of Calibration Experiment of Infrared Division-of-Focal-Plane Polarimeters

Xuancheng Peng1, Yuxiang Su1, Tie Hu1, Shichuan Wang1, Yuanlong Peng2, Jinkun Zheng3, Zhenyu Yang1, and Ming Zhao1、*
Author Affiliations
  • 1School of Optical and Electronic Information, Huazhong University of Science and Technology, Wuhan 430074, Hubei , China
  • 2State Grid Information & Telecommunication Branch, Beijing 100761, China
  • 3State Grid Jiangxi Electric Power Co., Ltd., Nanchang 330096, Jiangxi , China
  • show less
    Figures & Tables(15)
    Arrangement of the micro-polarizer array
    Schematic of NUC experiment setup
    Flow chart of NUC experiment
    Intensity images of different calibration light sources. (a) Non-uniform light source (laser light source with ring stripes); (b) uniform light source (blackbody)
    Images of DoLP and AoP of partial LPS calibrated by non-uniform light source. (a)(e) 0° LPS; (b)(f) 45° LPS; (c)(g) 90° LPS; (d)(h) 135° LPS
    Images of DoLP and AoP of partial LPS calibrated by uniform light source. (a)(e) 0° LPS; (b)(f) 45° LPS; (c)(g) 90° LPS; (d)(h) 135° LPS
    Polarization images of real scene calibrated by non-uniform light source. (a) Image of DoLP; (b) image of AoP
    Polarization relative error box plots of partial LPS after calibration with blackbody light source at different temperatures. (a) 0° LPS; (b) 45° LPS; (c) 90° LPS; (d) 135° LPS
    Mean polarization relative error as a function of number of super-pixels after calibration with blackbody at different temperatures
    Response curves of polarimeter pixels in different polarization directions before and after calibration. (a) Before calibration; (b) after calibration
    Near-field imaging calibration results. Images of (a) DoLP and (b) AoP before calibration; images of (c) DoLP and (d) AoP after calibration
    Far-field imaging calibration results. Images of (a) DoLP and (b) AoP before calibration; images of (c) DoLP and (d) AoP after calibration
    • Table 1. DNU of images calibrated by different light sources

      View table

      Table 1. DNU of images calibrated by different light sources

      LPSImage of DoLPImage of AoP
      Non-uniform light sourceUniform light sourceNon-uniform light sourceUniform light source
      0.09040.02461.25110.6095
      45°0.06070.02200.03180.0130
      90°0.09610.02100.02070.0066
      135°0.07430.02060.01310.0039
    • Table 2. Evaluation indices of near-field image quality

      View table

      Table 2. Evaluation indices of near-field image quality

      ConditionSDIEAG
      DoLPAoPDoLPAoPDoLPAoP
      Before calibration24.810546.32563.12384.55173.051516.6288
      After calibration60.683363.73203.79537.84116.933231.6898
    • Table 3. Evaluation indices of far-field image quality

      View table

      Table 3. Evaluation indices of far-field image quality

      ConditionSDIEAG
      DoLPAoPDoLPAoPDoLPAoP
      Before calibration20.38893.66706.11073.261210.06571.3306
      After calibration76.349133.10607.97096.260844.476311.2585
    Tools

    Get Citation

    Copy Citation Text

    Xuancheng Peng, Yuxiang Su, Tie Hu, Shichuan Wang, Yuanlong Peng, Jinkun Zheng, Zhenyu Yang, Ming Zhao. Error Analysis of Calibration Experiment of Infrared Division-of-Focal-Plane Polarimeters[J]. Acta Optica Sinica, 2024, 44(14): 1404001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Detectors

    Received: Mar. 6, 2024

    Accepted: Apr. 16, 2024

    Published Online: Jul. 19, 2024