Chinese Journal of Lasers, Volume. 50, Issue 9, 0907108(2023)

Point Cloud Registration Algorithm with Cross-Source and Low Overlapping Ratio for Pedicle Screw Fixation

Lijing Zhang1,2, Binbin Wang1,2, Wei Wang3, Bo Wu1,2、*, and Nan Zhang1,2
Author Affiliations
  • 1School of Biomedical Engineering, Capital Medical University, Beijing 100069, China
  • 2Beijing Key Laboratory of Fundamental Research on Biomechanics in Clinical Application, Capital Medical University, Beijing 100069, China
  • 3Department of Orthopedics, Xuanwu Hospital Capital Medical University, Beijing 100053, China
  • show less
    Figures & Tables(6)
    Framework for preoperative and intraoperative point clouds registration with cross-source and low overlapping based on farthest point sampling
    Sampling points of lumbar point cloud and their corresponding local regions. (a) Distribution of sampling points; (b) local regions corresponding to the sampling points
    Registration results of 001-L1, 001-L2, 002-L2, 002-L3, 003-L1, 004-L1, 005-L3, 006-L1, and 007-L1
    • Table 1. Overlapping ratio and location difference between intraoperative and preoperative point clouds

      View table

      Table 1. Overlapping ratio and location difference between intraoperative and preoperative point clouds

      DataExposed siteOverlapping ratio μ /%SourceRotation /(°)Translation /mm
      XYZXYZ
      001L11.7399.3546.6386.7974.86-28.8520.40
      001L21.69148.5348.47138.47114.08-64.99-103.52
      002L21.97-87.05-55.12106.0853.53-47.0829.48
      002L32.29109.4828.1979.30-55.677.2340.84
      003L11.35-76.5968.13-88.32-98.28-59.3626.07
      004L12.69104.6770.8697.17398.1034.1419.08
      005L31.8096.7042.0692.16216.39-44.5276.20
      006L11.6796.9422.4182.55-89.1145.3442.65
      007L11.6689.3958.1582.35-57.78-42.7338.71
    • Table 2. Comparison of preoperative and intraoperative point clouds coarse/fine registration errors and running time

      View table

      Table 2. Comparison of preoperative and intraoperative point clouds coarse/fine registration errors and running time

      DataExposed siteRegistration algorithmCoarse registrationFine registration
      Matching pairser /(°)er /mmTime /sMatching pairser /(°)er /mmTime /s
      001L1FPFH+ICP1523135.1326.5247.6511340.330.380.43
      SHOT+ICP1610019.9721.5455.7011330.710.660.40
      FPS+FPFH+ICP5729.279.80117.7811260.340.320.06
      001L2FPFH+ICP15
      SHOT+ICP1610536.4543.4358.979010.590.780.69
      FPS+FPFH+ICP26614.819.8192.309920.450.320.07
      002L2FPFH+ICP1518517.6225.7343.769210.931.210.32
      SHOT+ICP1610644.2164.3052.169150.640.850.56
      FPS+FPFH+ICP4864.074.76103.747430.340.430.06
      002L3FPFH+ICP152028.789.8644.4210141.000.410.35
      SHOT+ICP16
      FPS+FPFH+ICP3208.213.5489.5411420.250.85.0.08
      003L1FPFH+ICP1516319.8240.4947.0116830.450.990.54
      SHOT+ICP1615020.6235.2264.9016770.641.010.52
      FPS+FPFH+ICP3958.434.84115.1212930.830.440.06
      004L1FPFH+ICP15
      SHOT+ICP16
      FPS+FPFH+ICP2287.7517.8279.019900.310.470.05
      005L3FPFH+ICP1531125.5467.2345.5610540.260.930.27
      SHOT+ICP1624429.3948.0259.7710490.391.540.61
      FPS+FPFH+ICP8145.5411.8695.639200.370.640.04
      006L1FPFH+ICP15
      SHOT+ICP16
      FPS+FPFH+ICP5404.128.51113.548180.590.600.07
      007L1FPFH+ICP159522.711.1643.2514030.280.270.30
      SHOT+ICP168113.021.8157.6014070.430.580.31
      FPS+FPFH+ICP9942.793.00108.3413450.130.190.05
    • Table 3. Success rate comparison of preoperative and intraoperative point clouds registration

      View table

      Table 3. Success rate comparison of preoperative and intraoperative point clouds registration

      AlgorithmNumber of registration samplesNumber of successfully registered samplesSuccess rate /%
      FPFH+ICP159666.67
      SHOT+ICP169666.67
      FPS+FPFH+ICP99100
    Tools

    Get Citation

    Copy Citation Text

    Lijing Zhang, Binbin Wang, Wei Wang, Bo Wu, Nan Zhang. Point Cloud Registration Algorithm with Cross-Source and Low Overlapping Ratio for Pedicle Screw Fixation[J]. Chinese Journal of Lasers, 2023, 50(9): 0907108

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Biomedical Optical Imaging

    Received: Feb. 22, 2023

    Accepted: Mar. 3, 2023

    Published Online: Apr. 24, 2023

    The Author Email: Wu Bo (wubogo@ccmu.edu.cn)

    DOI:10.3788/CJL230561

    Topics