Laser & Optoelectronics Progress, Volume. 55, Issue 12, 122201(2018)

Development of Micro-Spectrophotometer and Its Application in Online Hexavalent Chromium Water Quality Monitoring Instrument

Xinchun Lu*, Yinghua Yuan, Yingqi Sun, Shudi Zhao, and Yonggang Luo
Author Affiliations
  • NARI Group Corporation/State Grid Electric Power Research Institute, Nanjing, Jiangsu 211000, China
  • show less
    Figures & Tables(12)
    Design diagram of spectrophotometer
    Structural diagram of optical system of micro-spectrophotometer
    Spot diagrams of system. (a) 200 nm; (b) 400 nm; (c) 550 nm; (d) 700 nm; (e) 900 nm
    Schematic of circuit design for micro-spectrophotometer
    Schematic of wavelength calibration experimental setup
    Spectrogram of calibration light source
    Spectrum of resolution testing
    Schematic of online hexavalent chromium water quality monitoring system
    Working curves of online hexavalent chromium water quality monitoring system. (a) Absorbance curves; (b) linear fitting curve
    • Table 1. Structural parameters of beam splitting system

      View table

      Table 1. Structural parameters of beam splitting system

      SurfacetypeRadius /mmThickness /mmTilt /(°)
      S40.0
      M180.035.08
      GInfinity37.518
      M2152.465.028
    • Table 2. Tolerance distributions of spectrophotometer

      View table

      Table 2. Tolerance distributions of spectrophotometer

      Tolerance typeM1M1 to GGG to M2M2
      Index tolerance Δn2C2C2C
      Abbe tolerance ΔV2C2C2C
      Fringe tolerance N (546.1 nm)11
      Surface irregularity tolerance ΔN (546.1 nm)0.50.5
      Surface decenter tolerance /mm0.050.050.05
      Surface tilt tolerance /(')111
      Interval tolerance Δd /mm0.050.05
      Compensator of back focus /mm0.1
    • Table 3. Wavelength calibration data

      View table

      Table 3. Wavelength calibration data

      FittingcoefficientPixelnumberStandardwavelength /nmCalibrationwavelength /nmRelativeerror /nm
      -46.066926801253.652253.6650.013
      0.38423091111365.015365.0290.014
      -1.1850444×10-51312435.833435.739-0.094
      -8.4414396×10-101633546.074546.1050.031
      1731579.066579.1500.084
      2087696.543696.534-0.009
      2296763.511763.439-0.072
      2581852.144852.1770.033
    Tools

    Get Citation

    Copy Citation Text

    Xinchun Lu, Yinghua Yuan, Yingqi Sun, Shudi Zhao, Yonggang Luo. Development of Micro-Spectrophotometer and Its Application in Online Hexavalent Chromium Water Quality Monitoring Instrument[J]. Laser & Optoelectronics Progress, 2018, 55(12): 122201

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Optical Design and Fabrication

    Received: Apr. 10, 2018

    Accepted: Jul. 5, 2018

    Published Online: Aug. 1, 2019

    The Author Email: Xinchun Lu (luxinchun@sgepri.sgcc.com.cn)

    DOI:10.3788/LOP55.122201

    Topics