Acta Optica Sinica, Volume. 45, Issue 8, 0812002(2025)

Correction Method of Stray Light in Spectrally Modulated Polarization Spectrometer

Yanghao Chu1,2,3, Feng Ji1, Mengfan Li2,3、*, Yadong Hu2,3、**, Zhenwei Qiu2,3, Jingjing Shi4, Xiangjing Wang2,3, Tao Wang2,3, Gaojun Chi2,3, Benyong Yang2,3, and Jin Hong2,3
Author Affiliations
  • 1Anhui Province Key Laboratory of Measuring Theory and Precision Instrument, School of Instrument Science and Opto-Electronics Engineering, Hefei University of Technology, Hefei 230009, Anhui, China
  • 2Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei 230031, Anhui, China
  • 3Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei 230031, Anhui, China
  • 4Deep Space Exploration Laboratory (Tiandu Laboratory), Hefei 230000, Anhui, China
  • show less
    Figures & Tables(19)
    Schematic of spectrally modulated polarization spectrometer system
    Modulated images of target light with different polarization states
    Stray light distributions. (a) Ray tracing results from LightTools; (b) measured results
    Spatial dimension stray light coefficient test. (a) With black band; (b) without black band
    Spatial dimension stray light coefficients at different fields of view. (a) S beam; (b) P beam
    Spectral dimension stray light coefficient test. (a) With notch filter; (b) without notch filter
    Spectral dimension stray light coefficients at different fields of view. (a) S beam; (b) P beam
    Impact of stray light on radiation measurement. (a)‒(d) Different stray light coefficients
    Impact of stray light on DoLP measurement. (a)‒(d) Different stray light coefficients
    Modulated spectral distributions on detector when DoLP of target light is 1 and AoLP is different
    Stray light distribution matrix measurement setup
    Experimental flow chart
    Stray light coefficient experimental correction results. (a) S beam; (b) P beam
    Radiation measurement correction results. (a)‒(c) Measured results; (d)‒(f) relative errors in radiation measurement
    DoLP measurement correction results. (a)‒(c) Measured results; (d)‒(f) absolute errors in DoLP measurement
    • Table 0. [in Chinese]

      View table

      Table 0. [in Chinese]

      NumberAoLP is 0°AoLP is 60°AoLP is 120°
      S beam wavelengthP beam wavelengthS beam wavelengthP beam wavelengthS beam wavelengthP beam wavelength
      13435.69428.22440.87425.51431.62430.13
      14444.14435.31449.31433.68438.96437.49
      15451.22443.76456.94440.49445.23446.22
      16460.76450.85465.93451.40453.67454.94
      17468.11459.57474.37457.12462.39462.30
      18478.46467.75483.64465.57471.92473.75
      19486.91478.11493.99473.20480.09480.02
      20496.71485.19503.80483.29491.27486.01
      21507.34495.29513.61492.29498.89492.02
      22517.69505.93525.05503.20510.88499.11
      23523.96517.38512.47521.78510.02
    • Table 1. Main path distributions of diffusion

      View table

      Table 1. Main path distributions of diffusion

      Number“Diffusion” path 1“Diffusion” path 2
      1L-PL-P
      2Detector surfaceFront surface of the protective window
      3Rear surface of the protective windowDetector surface
      4Detector surface
    • Table 2. Main path distributions of ghost

      View table

      Table 2. Main path distributions of ghost

      Number“Ghost” path 1“Ghost” path 1
      1L-PL-P
      2Detector surfaceDetector surface
      3Protective windowProtective window
      4SpectrometerSpectrometer
      5Rear surface of the filtersFilters
      6Front surface of the filtersTelescope
      7Rear surface of the filtersRear surface of PBS
      8SpectrometerTelescope
      9Protective windowFilters
      10Detector surfaceSpectrometer
      11Protective window
      12Detector surface
    • Table 3. Extremum wavelengths at 0° field of view

      View table

      Table 3. Extremum wavelengths at 0° field of view

      NumberAoLP is 0°AoLP is 60°AoLP is 120°
      S beam wavelengthP beam wavelengthS beam wavelengthP beam wavelengthS beam wavelengthP beam wavelength
      1339.22341.89340.03347.33343.32346.78
      2349.58347.60350.40353.52353.69356.80
      3355.60354.61360.99360.73362.64366.61
      4365.95361.00369.51367.16368.97372.61
      5377.14371.25380.68374.52375.51378.61
      6388.59382.15392.67380.25384.77384.34
      7395.13389.24399.21386.52390.50390.33
      8401.66395.24405.74392.24397.31396.88
      9407.65401.24412.28398.79403.57403.15
      10415.28407.24419.09405.33410.92409.69
      11421.54414.60425.08411.87416.64416.51
      12429.17420.87434.06418.42424.27423.60
    Tools

    Get Citation

    Copy Citation Text

    Yanghao Chu, Feng Ji, Mengfan Li, Yadong Hu, Zhenwei Qiu, Jingjing Shi, Xiangjing Wang, Tao Wang, Gaojun Chi, Benyong Yang, Jin Hong. Correction Method of Stray Light in Spectrally Modulated Polarization Spectrometer[J]. Acta Optica Sinica, 2025, 45(8): 0812002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Jan. 21, 2025

    Accepted: Feb. 26, 2025

    Published Online: Apr. 16, 2025

    The Author Email: Mengfan Li (mfli@aiofm.ac.cn), Yadong Hu (huyadong@aiofm.ac.cn)

    DOI:10.3788/AOS250529

    CSTR:32393.14.AOS250529

    Topics