Acta Photonica Sinica, Volume. 54, Issue 8, 0822002(2025)

Design of a Long-wave Infrared Medical Endoscope Optical System with 30° Viewing Angle

Duoyao WANG1,2,3, Wenlong LI1,2、**, Qianlong XIE3,4, Chengrun LI5, Hongxu YUE5, Rui YAN5, and Depeng KONG1,2、*
Author Affiliations
  • 1State Key Laboratory of Ultrafast Optical Science and Technology, Xi'an Institute of Optics and Precision Mechanics, Chinese Academy of Sciences, Xi'an 710119, China
  • 2Photonics Functional Materials and Devices Research Laboratory, Xi'an Institute of Optics and Precision Mechanics, Chinese Academy of Sciences, Xi'an 710119, China
  • 3School of Optoelectronics, University of Chinese Academy of Sciences, Beijing 101408, China
  • 4Key Laboratory of Spectral Imaging Technology, Xi'an Institute of Optics and Precision Mechanics of CAS, Xi'an 710119,China
  • 5Department of Thoracic Surgery, Chinese People's Liberation Army General Hospital, Beijing 100853, China
  • show less
    Figures & Tables(13)
    Objective lens module
    Relay module
    Coupled mirror module
    Structure of endoscope system
    MTF curve
    Spot diagram
    Relative illumination
    Field curvature and distortion diagram
    MTF curves under different temperature conditions
    MTF curves under different temperature conditions after thermal compensation
    • Table 1. Design parameters of the endoscope system

      View table
      View in Article

      Table 1. Design parameters of the endoscope system

      ParameterValue
      F/#1.39
      View angle/(°)30°
      Working wavelength/μm8~14
      Field of view/(°)±19.4°
      Working distance/mm100
      Insertion length/mm≥200
      Total length/mm≤300
      Diameter/mm≤9
      Back focal length/mm≥25
      NETD/mK45
    • Table 2. Tolerance limit value

      View table
      View in Article

      Table 2. Tolerance limit value

      ItemValue
      TFRN/rings3
      TTHI/mm0.015
      TSDX/mm0.015
      TSDY/mm0.015
      TEDX/mm0.015
      TEDY/mm0.015
      TSTX/(°)0.015
      TSDY/(°)0.015
      TETX/(°)0.015
      TETY/(°)0.015
      Refractive index0.000 3
      Abbe-number0.5
    • Table 3. Monte Carlo analysis results

      View table
      View in Article

      Table 3. Monte Carlo analysis results

      Percentage/%The value of MTF at 41.7 lp/mm
      >98%0.198 216 83
      >90%0.209 607 76
      >80%0.214 554 22
      >50%0.222 371 36
      >20%0.228 607 66
    Tools

    Get Citation

    Copy Citation Text

    Duoyao WANG, Wenlong LI, Qianlong XIE, Chengrun LI, Hongxu YUE, Rui YAN, Depeng KONG. Design of a Long-wave Infrared Medical Endoscope Optical System with 30° Viewing Angle[J]. Acta Photonica Sinica, 2025, 54(8): 0822002