Spectroscopy and Spectral Analysis, Volume. 40, Issue 12, 3806(2020)

Application of High Pressure Pelletised Sample and LIBS in the Determination of Rare Earth Elements in Soil Samples

Jin-li XU1,1、*, Meng-ying HU1,1, Peng-peng ZHANG1,1, Xia XING1,1, Jin-feng BAI1,1, and Qin ZHANG1,1
Author Affiliations
  • 1[in Chinese]
  • 11. Institute of Geophysical and Geochemical Exploration, Chinese Academy of Geological Sciences, Langfang 065000, China
  • show less
    Figures & Tables(7)
    Effection of sample preparation pressure on the precision of measurement results
    Characteristic spectrum of GSS21 and GSS24 under different pressures (1 000~2 500 kN)
    Effection of laser ablation mode on the measurement results of GSS7 and GSS20
    Effection of standard curve of GSS1—10 on measurement results
    Effection of standard curve of GSS1—10 on Prision
    • Table 1. Standard curve of rare earth elements

      View table
      View in Article

      Table 1. Standard curve of rare earth elements

      ElementUnivariate equationCorrelation
      coefficient(R)
      P(n=10)
      Test
      Yy=0.909 4x+2.063 20.994 1<0.000 1
      Lay=0.937 9x+3.994 90.996 4<0.000 1
      Cey=0.941 2x+8.651 80.999 5<0.000 1
      Pry=0.941 3x+1.183 80.997 1<0.000 1
      Ndy=0.941 2x+4.406 50.997 3<0.000 1
      Smy=0.934 5x+0.548 70.997 4<0.000 1
      Euy=0.933 5x+0.103 10.997 8<0.000 1
      Gdy=0.932 1x+0.352 70.999 0<0.000 1
      Tby=0.914 5x+0.066 80.998 3<0.000 1
      Dyy=0.974 7x+0.112 10.998 0<0.000 1
      Hoy=0.900 2x+0.0870.998 4<0.000 1
      Ery=0.914 1x+0.200 90.997 9<0.000 1
      Tmy=0.908 7x+0.036 70.999 3<0.000 1
      Yby=0.916 1x+0.201 40.997 7<0.000 1
      Luy=0.908 6x+0.035 30.994 9<0.000 1
      校准曲线的标准物质: GBW07401, GBW07402, GBW07403, GBW07404, GBW07405, GBW07406, GBW07407, GBW07408, GBW07423, GBW07424
    • Table 2. Detection limit

      View table
      View in Article

      Table 2. Detection limit

      ElementBlankSoilReferences[15]
      detection limit
      /(mg·kg-1)
      Ave
      (n=12)
      sdetection limit
      /(mg·kg-1)
      Ave
      (n=12)
      sdetection limit
      /(mg·kg-1)
      Y13.21.574.7216.60.732.191
      La6.912.888.6534.50.461.381
      Ce13.36.6619.9759.41.243.732
      Pr3.190.571.707.700.100.290.1
      Nd10.92.367.0927.20.441.330.1
      Sm2.130.401.204.410.110.330.1
      Eu0.930.020.061.200.000.010.1
      Gd2.600.341.033.770.110.340.1
      Tb0.340.060.180.550.020.070.1
      Dy2.170.331.002.860.150.440.1
      Ho0.460.060.180.610.030.080.1
      Er1.260.180.531.710.080.240.1
      Tm0.210.030.090.270.010.040.1
      Yb1.190.210.621.660.100.290.1
      Lu0.180.030.100.260.010.040.1
    Tools

    Get Citation

    Copy Citation Text

    Jin-li XU, Meng-ying HU, Peng-peng ZHANG, Xia XING, Jin-feng BAI, Qin ZHANG. Application of High Pressure Pelletised Sample and LIBS in the Determination of Rare Earth Elements in Soil Samples[J]. Spectroscopy and Spectral Analysis, 2020, 40(12): 3806

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Dec. 11, 2019

    Accepted: --

    Published Online: Jun. 18, 2021

    The Author Email: Jin-li XU (80368070@qq.com)

    DOI:10.3964/j.issn.1000-0593(2020)12-3806-06

    Topics