Laser & Optoelectronics Progress, Volume. 60, Issue 20, 2011002(2023)

Design of a Capsule Endoscope Optical System with a Large Depth of Field

Jiegen Yang, Feng Hu, Wei Ding, and Liner Zou*
Author Affiliations
  • School of Physics and Materials Science, Nanchang University, Nanchang 330031, Jiangxi , China
  • show less
    Figures & Tables(15)
    Schematic of capsule endoscope. (a) Overall appearance drawing; (b) internal structure diagram
    Optical system structure diagram
    MTF curve at a working distance of 12 mm
    Spot diagram at a working distance of 12 mm
    Optical system structure diagrams and spot diagrams at different working distance. (a) 7 mm; (b) 20 mm; (c) 50 mm; (d) 100 mm
    Field curved and distortion. (a) Field curved; (b) distortion
    Simulated images. (a) 12th unit image of JB/T 9328-1999 A3 resolution board; (b) simulated image of the center field; (c) simulated image of the edge field
    Simulated images. (a) 9th unit image; (b) simulated image of the center field; (c) simulated image of the edge field; (d) 8th unit image; (e) simulated image of the center field; (f) simulated image of the edge field
    MTF curves at different working distance. (a) 7 mm; (b) 20 mm; (c) 50 mm; (d) 100 mm
    • Table 1. Optimization results with different F numbers

      View table

      Table 1. Optimization results with different F numbers

      F/#MTF(125 lp/mm)Depth of field /mm
      4>0.57‒30
      5>0.357‒100
      6>0.256‒100
    • Table 2. Parameters of imaging chip

      View table

      Table 2. Parameters of imaging chip

      Optical size /inchPixel size /(μm×μm)Effective area /(μm×μm)Resolution
      1/91.4×1.41820×10331280×720
    • Table 3. System design parameters

      View table

      Table 3. System design parameters

      ParameterData
      Depth of field /mm7‒100
      Field of view /(°)100
      F/#5
      MTF125 line pairs>0.35
    • Table 4. Tolerance values of the systems

      View table

      Table 4. Tolerance values of the systems

      ItemValue
      TFRN /fringe2
      TTHI /mm0.006
      TSDY /mm0.005
      TSTY /(°)0.0167
      TEDY /mm0.005
      TETY /(°)0.005
    • Table 5. Monte Carlo results of the tolerance analysis

      View table

      Table 5. Monte Carlo results of the tolerance analysis

      Percentage /%The value of MTF at 172 lp/mm
      90> 0.20801540
      80> 0.22677880
      50> 0.26245544
      20> 0.27872328
      10> 0.28508444
    • Table 6. Most sensitive items of tolerance analysis

      View table

      Table 6. Most sensitive items of tolerance analysis

      TypeSurfaceValueCriterionChange
      TEDY99-0.0050.25053151-0.05128430
      TEDY990.0050.25053151-0.05128430
      TEDY10100.0050.25432082-0.04749499
      TEDY1010-0.0050.25432082-0.04749499
      TEDY1111-0.0050.25969552-0.04212030
      TEDY11110.0050.25969552-0.04212030
      TEDY880.0050.26965158-0.03216423
      TEDY88-0.0050.26965158-0.03216423
      TFRN10-20.27928139-0.02253443
      TEDY660.0050.28033010-0.02148572
    Tools

    Get Citation

    Copy Citation Text

    Jiegen Yang, Feng Hu, Wei Ding, Liner Zou. Design of a Capsule Endoscope Optical System with a Large Depth of Field[J]. Laser & Optoelectronics Progress, 2023, 60(20): 2011002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Imaging Systems

    Received: Sep. 21, 2022

    Accepted: Dec. 12, 2022

    Published Online: Oct. 13, 2023

    The Author Email: Liner Zou (Linerzou@ncu.edu.cn)

    DOI:10.3788/LOP222593

    Topics