Acta Optica Sinica, Volume. 42, Issue 23, 2322001(2022)

Optical System Design of Full-Spectrum Hyperspectral Imaging Spectrometer

Jian Ma1,2, Junqiang Zhang1,3、*, Congjun Wu1、**, Xinquan Li1,2, Tianjiao Lu1,2, and Bin Yang3
Author Affiliations
  • 1Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130031, Jilin , China
  • 2University of Chinese Academy of Sciences, Beijing 101400, China
  • 3Yusense Information Technology and Equipment (Qingdao) Inc., Qingdao 260000, Shandong , China
  • show less
    Figures & Tables(13)
    Design scheme of optical system
    Coaxial three-mirror system
    Optical path diagram of front telescopic system. (a) Overall optical path; (b) optical path at image plane
    MTF curves of each spectral segment. (a) VNIR spectrum ;(b) SWIR spectrum ;(c) MWIR spectrum;(d) LWIR spectrum
    Optical path diagram of Offner spectral imaging system with convex grating. (a) Incident optical path ; (b) outgoing optical path
    Optical path diagram of Offner spectrometer in each spectral band. (a) VNIR; (b) SWIR; (c) MWIR; (d) LWIR
    Optical structure diagram of whole system
    MTF curve of center wavelength of each spectral band in whole system. (a) 0.7 μm; (b) 1.7 μm; (c) 4.0 μm; (d) 10.2 μm
    Cross-sectional view of point spread function of each spectral band. (a) VNIR; (b) SWIR; (c) MWIR; (d) LWIR
    • Table 1. Technical specifications of optical system

      View table

      Table 1. Technical specifications of optical system

      SpecificationVNIRSWIRMWIRLWIR
      Spectral range /μm0.4-1.01.0-2.53.0-5.08.0-12.5
      Field of view /(°)2.8642.8642.8642.864
      Focal length /mm400400360244

      Detector pixel size /

      (μm×μm)

      20×2020×2030×3030×30
      Number of spatial dimension pixels10241024640640
      Spectral resolution /nm51050100
      Pixel resolution /m30305074
    • Table 2. Design parameters of multiple configurations

      View table

      Table 2. Design parameters of multiple configurations

      Design parameterConfiguration 1Configuration 2Configuration 3Configuration 4
      Off-axis field of view /(°)-9.3-9.3-9.0-8.7
      FOV No.1123
      SpectrumVNIRSWIRLWIRMWIR
      Spectral range /μm0.4-1.01.0-2.58.0-12.53.0-5.0
    • Table 3. Initial structural parameters of Offner spectrometer in each spectral band

      View table

      Table 3. Initial structural parameters of Offner spectrometer in each spectral band

      ParameterVNIRSWIRMWIRLWIR
      RG /mm66.6766.6760.0050.00
      Slit width /μm20.0020.0033.3049.18
      Slit length /mm20202020
      Magnification1.001.000.900.61
      Spectral broadening /mm2.403.001.201.35
    • Table 4. MTF of different wavelengths at Nyquist frequency for each spectral band spectrometer

      View table

      Table 4. MTF of different wavelengths at Nyquist frequency for each spectral band spectrometer

      SpectrumWavelength /μmMTF
      VNIR0.40.84
      0.70.74
      1.00.70
      SWIR1.00.65
      1.70.70
      2.50.65
      MWIR3.00.80
      4.00.77
      5.00.71
      LWIR8.00.66
      10.20.60
      12.50.55
    Tools

    Get Citation

    Copy Citation Text

    Jian Ma, Junqiang Zhang, Congjun Wu, Xinquan Li, Tianjiao Lu, Bin Yang. Optical System Design of Full-Spectrum Hyperspectral Imaging Spectrometer[J]. Acta Optica Sinica, 2022, 42(23): 2322001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Optical Design and Fabrication

    Received: Apr. 1, 2022

    Accepted: Jun. 8, 2022

    Published Online: Dec. 14, 2022

    The Author Email: Zhang Junqiang (zhangjq@ciomp.ac.cn), Wu Congjun (wucongjun@ciomp.ac.cn)

    DOI:10.3788/AOS202242.2322001

    Topics