Acta Optica Sinica, Volume. 43, Issue 12, 1212004(2023)

Mueller Matrix Measurement of Off-Axis Three-Mirror Telescope Objective

Xinxin Zhao1,2,3、**, Maoxin Song2,3、*, Zhilong Xu2,3, Dapeng Kuang2,3, Guangfeng Xiang2,3, and Jin Hong1,2,3、***
Author Affiliations
  • 1School of Environmental Science and Optoelectronic Technology, University of Science and Technology of China, Hefei 230026, Anhui, China
  • 2Anhui Institute of Optics and Fine Mechanics, Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei 230031, Anhui, China
  • 3Key Laboratory of Optical Calibration and Characterization, Chinese Academy of Sciences, Hefei 230031, Anhui, China
  • show less
    Figures & Tables(9)
    Schematic of Mueller matrix dual-rotating retarder polarimeter
    Diagram of system parameters calibration experiment device
    Relationship of retardance and temperature fitted by linear least square method. (a) Waveplate 1; (b) waveplate 2
    Schematic of off-axis three-mirror objective Mueller matrix measurement system
    Diagram of objective Mueller matrix measurement device
    m12 and m13 fitted by nonlinear least square method
    When input 1001 Gaussian noise, variation of measurement error of polarization under the condition of p=1.0, 0.8, 0.5, 0.3, 0.1. (a) Measurement error of Dp; (b) measurement error of Dlp; (c) measurement error of Dcp
    • Table 1. Device parameters of measurement system

      View table

      Table 1. Device parameters of measurement system

      ParameterValue
      Stability of light source /h6×10-4
      Extinction ratio of Glan-Taylor prism and polarizer> 105
      Accuracy of the rotator /(°)0.005
      Retardance accuracy of 1/4 achromatic waveplate @400-700 nmλ/100
      Linearity of detector /%±0.5
      Temperature measurement accuracy of PT100 sensor /℃±0.1
      Angular positioning accuracy of theodolite /(°)0.005
    • Table 2. Measurement results of system parameters

      View table

      Table 2. Measurement results of system parameters

      Parameterε3 /(°)ε4 /(°)ε5 /(°)k1 /[(°)·℃-1k2 /[(°)·℃-1b1 /(°)b2 /(°)
      Value0.26640.48580.47890.08620.0777-4.1563-4.9523
    Tools

    Get Citation

    Copy Citation Text

    Xinxin Zhao, Maoxin Song, Zhilong Xu, Dapeng Kuang, Guangfeng Xiang, Jin Hong. Mueller Matrix Measurement of Off-Axis Three-Mirror Telescope Objective[J]. Acta Optica Sinica, 2023, 43(12): 1212004

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Oct. 25, 2022

    Accepted: Dec. 1, 2022

    Published Online: Jun. 20, 2023

    The Author Email: Zhao Xinxin (heyuee@mail.ustc.edu.cn), Song Maoxin (smx0369@163.com), Hong Jin (hongjin@aiofm.ac.cn)

    DOI:10.3788/AOS221873

    Topics