Chinese Journal of Lasers, Volume. 51, Issue 13, 1306005(2024)

Research on Sensing Cardiac Cycles Using a Microbend Fiber Sensor

Zhiguo Jiang1,4, Jianing Ma2, Shuxia Qian3, Chaochao Wang1,4, and Xianchao Zhang1,4、*
Author Affiliations
  • 1Key Laboratory of Medical Electronics and Digital Health of Zhejiang Province, Jiaxing University, Jiaxing 314001, Zhejiang , China
  • 2School of Information and Communication Engineering, Beijing University of Posts and Telecommunications, Beijing 100876, China
  • 3Department of Neurology, Second Affiliated Hospital of Jiaxing University, Jiaxing 314000, Zhejiang , China
  • 4Engineering Research Center of Intelligent Human Health Situation Awareness of Zhejiang Province, Jiaxing University, Jiaxing 314001, Zhejiang , China
  • show less
    Figures & Tables(13)
    BCG signal and its characteristic peaks
    Microbend fiber optic vital sign sensor. (a) Sensor structure diagram; (b) photo of sensor
    Cardiac cycle detection method based on microbend fiber optic sensing. (a) Method schematic diagram; (b) photo during the experiment
    Motion artifacts caused by body movement
    Weak heart and lung vibration signal collected in the experiment. (a) Time-domain waveform; (b) time frequency scale map based on wavelet transform
    IMF components obtained after 4-layer VMD decomposition
    BCG waveform obtained after SPA-IVMD joint denoising
    J-peak adaptive positioning process. (a) Preselected J-peak; (b) second peak after preselected J-peak; (c) results filtered by amplitude threshold; (d) results filtered by peak time interval
    J-wave peak localization of BCG
    Comparison of cardiac cycles obtained from ECG and BCG respectively
    • Table 1. Center frequency of each IMF component under different decomposition levels

      View table

      Table 1. Center frequency of each IMF component under different decomposition levels

      Decomposition levelIMF1IMF2IMF3IMF4IMF5
      13.407
      22.5226.044
      32.4345.46715.129
      42.2164.5997.42316.602
      52.2064.5327.00814.28372.211
    • Table 2. Calculation of average heart rate

      View table

      Table 2. Calculation of average heart rate

      SubjectBCG /(beat·min-1ECG /(beat·min-1Error /%
      166.40 (66 peaks)66.33 (66 peaks)0.106
      253.19 (53 peaks)53.10 (53 peaks)0.169
      357.86 (56 peaks)57.81 (58 peaks)0.086
      474.27 (72 peaks)73.19 (73 peaks)1.476
      570.16 (70 peaks)70.07 (70 peaks)0.128
    • Table 3. Average heart rate under different algorithms

      View table

      Table 3. Average heart rate under different algorithms

      Method

      ECG /

      (beat·min-1

      BCG /

      (beat·min-1

      Average

      error /%

      STE

      66.0439

      70.61216.92
      TM64.76121.94
      CA64.11462.92
      Proposed66.49810.69
    Tools

    Get Citation

    Copy Citation Text

    Zhiguo Jiang, Jianing Ma, Shuxia Qian, Chaochao Wang, Xianchao Zhang. Research on Sensing Cardiac Cycles Using a Microbend Fiber Sensor[J]. Chinese Journal of Lasers, 2024, 51(13): 1306005

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Fiber optics and optical communication

    Received: Dec. 11, 2023

    Accepted: Jan. 25, 2024

    Published Online: Jul. 2, 2024

    The Author Email: Xianchao Zhang (zhangxianchao@zjxu.edu.cn)

    DOI:10.3788/CJL231506

    CSTR:32183.14.CJL231506