Journal of Infrared and Millimeter Waves, Volume. 42, Issue 3, 369(2023)

Research on nonlinear correction algorithm of Fourier transform spectrometer based on responsivity correction

Yao-Pu ZOU1,2, Xuan FENG3、*, Chang-Pei HAN1,2、**, Li-Bing LI1,2, Xiao LIANG2,4, Jing QIAN1,2, and Xiu-Ju LI1,2
Author Affiliations
  • 1Key Laboratory of Infrared System Detection and Imaging Technology, Shanghai 200083, China
  • 2Shanghai Institute of Technical Physics, Chinese Academy of Sciences, Shanghai 200083, China
  • 3Key Laboratory of Space Utilization, Technology and Engineering center for Space Utilization, Chinese Academy of Sciences, Beijing 100094, China
  • 4University of Chinese Academy of Sciences, Beijing 100049, China
  • show less
    Figures & Tables(14)
    Interferogram nonlinear response schematic
    (a)The schematic diagram of nonlinear interferogram, (b)spectral out-of-band distortion due to quadratic nonlinearity
    Variation of phase angle of different blackbody brightness temperature responsivity
    Variation of blackbody spectral responsivity phase angle for 5 months in orbit
    Simulation of approximate linear relationship between spectral responsivity and spectral value
    Long-wave full-band responsivity correction coefficient of TVAC1-TVAC5 fitted value and average value
    The linear relationship between the target spectral value and the spectral responsivity of TVAC1-TVAC5
    Brightness temperature of radiometric calibration comparison before and after nonlinear correction
    Brightness temperature deviations comparison before and after nonlinear correction
    Brightness temperature standard deviation comparison before and after nonlinear correction
    Comparison of calibration deviations before and after nonlinear correction for all pixels in the range of 180-320 K of brightness temperature of the observation target: (a) 850 cm-1, (b) 950 cm-1
    The average deviation of calibration after nonlinear correction of all pixels in the range of 180-320 K brightness temperature of the observation target
    Observation target brightness temperature 280 K TVAC1-TVAC5 calibration average deviation after nonlinear correction
    • Table 1. TVAC tests of GIIRS radiometric calibration

      View table
      View in Article

      Table 1. TVAC tests of GIIRS radiometric calibration

      测温点

      名称

      不同工况下仪器部分结构测温(摄氏度)
      TVAC1TVAC2TVAC3TVAC4TVAC5
      扫描镜261313540
      北冷板1000-1020
      隔热罩2-810-1112
      散热北板-151212-2019
      散热南板-10-10-10-4020
    Tools

    Get Citation

    Copy Citation Text

    Yao-Pu ZOU, Xuan FENG, Chang-Pei HAN, Li-Bing LI, Xiao LIANG, Jing QIAN, Xiu-Ju LI. Research on nonlinear correction algorithm of Fourier transform spectrometer based on responsivity correction[J]. Journal of Infrared and Millimeter Waves, 2023, 42(3): 369

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Oct. 26, 2022

    Accepted: --

    Published Online: Jul. 5, 2023

    The Author Email: Xuan FENG (fengxuan@csu.ac.cn), Chang-Pei HAN (changpei_han@mail.sitp.ac.cn)

    DOI:10.11972/j.issn.1001-9014.2023.03.012

    Topics