Spectroscopy and Spectral Analysis, Volume. 42, Issue 1, 197(2022)

Detection of C Element in Alloy Steel by Double Pulse Laser Induced Breakdown Spectroscopy With a Multivariable GA-BP-ANN

Feng-ping YU*, Jing-jun LIN1; *;, Xiao-mei LIN1; 3; *;, and Lei LI1; 2; *;
Author Affiliations
  • 1. Department of Electronics and Electrical Engineering, Changchun University of Technology, Changchun 130012, China
  • show less
    Figures & Tables(9)
    Schematic diagram of the experimental setup
    (a)The fitting curve of C content (%) with its spectral line intensity (b) the fitting curve of C content (%) with C / Fe spectral lines intensity ratios
    The DP-LIBS spectrum of alloy steel and selected characteristic analysis spectrum lines
    The correlation curve between predicted content and certified content of element C by internal calibration, multivariable GA-BP-ANN of and univariate BP-ANN
    • Table 1. Reference concentration (%) of element C and Fe in the samples

      View table
      View in Article

      Table 1. Reference concentration (%) of element C and Fe in the samples

      12345678910
      C0.0810.0990.1220.2220.3580.4180.5800.6640.7470.881
      Fe98.77398.22697.23694.11697.23896.91496.48998.02194.88596.492
    • Table 2. The relative error (RE%) of internal calibration in predicted content of element C

      View table
      View in Article

      Table 2. The relative error (RE%) of internal calibration in predicted content of element C

      No.12345678910
      RE/%49.3853.5418.038.1110.0610.7710.356.338.577.26
    • Table 3. The relative error (RE%) of univariate BP-ANN calibration in predicted content of element C

      View table
      View in Article

      Table 3. The relative error (RE%) of univariate BP-ANN calibration in predicted content of element C

      No.12345678910
      RE/%64.2033.3349.1810.3615.0816.276.553.311.345.56
    • Table 4. The relative error (RE%) of multivariable GA-BP-ANN calibration in predicted content of element C

      View table
      View in Article

      Table 4. The relative error (RE%) of multivariable GA-BP-ANN calibration in predicted content of element C

      No.12345678910
      RE/%98.76527.276.568.117.8211.006.902.561.472.95
    • Table 5. The average of relative error (ARE%) and relative standard deviation (ARSD%) by internal calibration, multivariable GA-BP-ANN of and univariate BP-ANN predicted content of element C

      View table
      View in Article

      Table 5. The average of relative error (ARE%) and relative standard deviation (ARSD%) by internal calibration, multivariable GA-BP-ANN of and univariate BP-ANN predicted content of element C

      Internal
      calibration
      BP-ANNGA-BP-ANN
      ARE/%14.7814.758.29
      ARSD/%12.38511.1238.818
    Tools

    Get Citation

    Copy Citation Text

    Feng-ping YU, Jing-jun LIN, Xiao-mei LIN, Lei LI. Detection of C Element in Alloy Steel by Double Pulse Laser Induced Breakdown Spectroscopy With a Multivariable GA-BP-ANN[J]. Spectroscopy and Spectral Analysis, 2022, 42(1): 197

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Dec. 3, 2020

    Accepted: --

    Published Online: Mar. 31, 2022

    The Author Email: Feng-ping YU (1171207294@qq.com)

    DOI:10.3964/j.issn.1000-0593(2022)01-0197-06

    Topics