Chinese Journal of Lasers, Volume. 49, Issue 21, 2110002(2022)

Dewar Packaging Technology of Multiband Long-Wave Infrared Focal Plane Array Detectors for Cryogenic Optics

Junlin Chen1,2, Xiaokun Wang1,2、*, Haiyong Zhu1,2, Zhijiang Zeng1,2, Yaran Li2, Shaobo Luo1,2, Xi Wang1,2, and Chun Lin1,2
Author Affiliations
  • 1State Key Laboratories of Transducer Technology, Shanghai Institute of Technical Physics, Chinese Acadamy of Sciences, Shanghai 200083, China
  • 2Shanghai Institute of Technical Physics, Chinese Academy of Sciences, Shanghai 200083, China
  • show less
    Figures & Tables(12)
    Schematic of mercury cadmium telluride (MCT)
    Schematic of multi-band MCT stitching
    Structure of Dewar
    Schematic of filter structure
    Simulation analysis of the influence of filter position on Dewar’s optical characteristics. (a) Optical and mechanical structure of proposed system; (b) non-uniformity of detector; (c) stray light analysis of M1 module; (d) stray light analysis of M2 module
    Schematic of heat leakage measurement of bellows
    Temperature related to infrared calibration test
    Cold platform structure for isolating coupling stress
    Interference coupling stress test
    Photo of Dewar
    • Table 1. Measurement of thermal resistance of bellows

      View table

      Table 1. Measurement of thermal resistance of bellows

      No.Q /WT1 /KT2 /KΔT /K
      1074.45153.579.05
      20.10177.6193115.4
    • Table 2. Main performance indexes of infrared focal plane array Dewar

      View table

      Table 2. Main performance indexes of infrared focal plane array Dewar

      ItemContentItemContent
      Temperature /K55Operative pixel percent99.5%
      Wave band /μmB1: 11.5-12.5 μmResponse inhomogeneity10%
      B2: 8.0-10.5 μm
      B3: 0.3-11.3 μm
      Temperature difference /K±0.4Maximum size /(mm×mm×mm)115×115×126
      Pixed number2000×12Adapter coolerSpecial
      Pixel size /(μm×μm)30×30Heat load /mW1350
      PackagingDewarAssembly mass/g1750
    Tools

    Get Citation

    Copy Citation Text

    Junlin Chen, Xiaokun Wang, Haiyong Zhu, Zhijiang Zeng, Yaran Li, Shaobo Luo, Xi Wang, Chun Lin. Dewar Packaging Technology of Multiband Long-Wave Infrared Focal Plane Array Detectors for Cryogenic Optics[J]. Chinese Journal of Lasers, 2022, 49(21): 2110002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: remote sensing and sensor

    Received: Jan. 5, 2022

    Accepted: Mar. 8, 2022

    Published Online: Nov. 9, 2022

    The Author Email: Wang Xiaokun (wxk-sitp@mail.sitp.ac.cn)

    DOI:10.3788/CJL202249.2110002

    Topics