Influence of scanning velocity on bovine shank bone ablation with pulsed CO2 laser
The influence of scanning speed on hard bone tissue ablation is studied with a 10.6-\mum laser. The groove morphology and the thermal damage created in bovine shank bone by pulsed CO2 laser are examined as a function of incident fluence by optical microscope following standard histological processing. The results show that ablation groove width, depth and ablation volume, as well as the zone of thermal injury, increase gradually with incident fluence. As compared to the result for high scanning speed, the lower scanning speed always produces larger ablation volume but thicker zone of thermal injury. It is evident that scanning speed plays an important role in the ablation process. In clinical applications, it is important to select appropriate scanning speed to obtain both high ablation rates and minimal thermal injury.